FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 16414137 · Received February 21, 2023

Report

Report Number
1221359-2023-00384
Event Type
Malfunction
Date Received
February 21, 2023
Date of Event
February 15, 2023
Report Date
April 26, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 210543 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 210543 AND DEVICE PART NUMBER 195-430WL / LOT 207460. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 210543 SHOWED THAT THE COMPLAINT RATE IS (B)(4). BASED ON THE ABOVE SUMMARY, THE INVESTIGATION IS DEEMED COMPLETE. THE PRODUCT WILL CONTINUE TO BE MONITORED AND TRACKED. H3 OTHER TEXT : SINGLE USE DEVICE DISCARDED.

Additional Manufacturer Narrative · 0

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. SINGLE USE DEVICE DISCARDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED TWO (2) FALSE NEGATIVE RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2023. THIS MFR REPORT IS FOR TEST TWO (2) OF TWO (2). THE PATIENT PERFORMED AN ADDITIONAL TEST ON THE SAME DAY WITH PCR (PLATFORM UNKNOWN) AND TESTED POSITIVE. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED TWO (2) FALSE NEGATIVE RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2023. THIS MFR REPORT IS FOR TEST TWO (2) OF TWO (2). THE PATIENT PERFORMED AN ADDITIONAL TEST ON THE SAME DAY WITH PCR (PLATFORM UNKNOWN) AND TESTED POSITIVE. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1598441 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 210543 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 Female