FDA Adverse Event Malfunction Summary report: N

ADAC TRANSCAM

MDR report key: 1641320 · Received March 9, 2010

Report

Report Number
1641320
Event Type
Malfunction
Date Received
March 9, 2010
Date of Event
February 5, 2010
Report Date
March 9, 2010
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
IYX
Product Problem
Yes
Report Source
User Facility report
Reporter Location
FL, US

Narratives

Description of Event or Problem · 1

TRANSCAM NUCLEAR MEDICINE CAMERA WAS BEING PREPPED TO BE USED. DURING PREP, THE TECH SAID THE DETECTOR SUDDENLY CAME FORWARD TOWARDS HIM AND FELL TO THE FLOOR. THREE OF THE RIGHT BOLTS HOLDING THE DETECTOR TO THE YOKE WERE BENT, BOLTS APPEARS TO HAVE PULLED THROUGH THREADS IN THE DETECTOR. NO HARM TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADAC TRANSCAM NUCLEAR MEDICINE, GAMMA CAMERA IYX PHILIPS MEDICAL SYSTEMS PN 5200-2338 *

Patients

Seq Age Sex Outcome Treatment
1 50 YR