FDA Adverse Event
Malfunction
Summary report: N
ADAC TRANSCAM
MDR report key: 1641320
·
Received March 9, 2010
Report
- Report Number
- 1641320
- Event Type
- Malfunction
- Date Received
- March 9, 2010
- Date of Event
- February 5, 2010
- Report Date
- March 9, 2010
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- IYX
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- FL, US
Narratives
Description of Event or Problem · 1
TRANSCAM NUCLEAR MEDICINE CAMERA WAS BEING PREPPED TO BE USED. DURING PREP, THE TECH SAID THE DETECTOR SUDDENLY CAME FORWARD TOWARDS HIM AND FELL TO THE FLOOR. THREE OF THE RIGHT BOLTS HOLDING THE DETECTOR TO THE YOKE WERE BENT, BOLTS APPEARS TO HAVE PULLED THROUGH THREADS IN THE DETECTOR. NO HARM TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADAC TRANSCAM | NUCLEAR MEDICINE, GAMMA CAMERA | IYX | PHILIPS MEDICAL SYSTEMS | PN 5200-2338 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |