RELION INSULIN SYRINGE 3/10ML 29G X 8MM
Report
- Report Number
- 1920898-2023-00060
- Event Type
- Malfunction
- Date Received
- February 21, 2023
- Date of Event
- January 26, 2023
- Report Date
- June 9, 2023
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- UDI-DI
- 00681131311786
- PMA / PMN Number
- K212499
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- 003
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES, D10: RETURNED TO MANUFACTURER ON: 24-FEB-2023. H6: INVESTIGATION SUMMARY CUSTOMER RETURNED (4) SYRINGE 0.3ML 31GA 8MM IN AN OPEN POLY BAG FROM LOT# 2213837. IT WAS REPORTED BY THE CONSUMER THAT LIQUID COMES OUT FROM SYRINGE WHEN DEPRESSING PLUNGER ROD. THE 4 RETURN SAMPLES FROM THE OPEN POLY BAG WERE TESTED AND SMALL CLEAR DROPLET OF MATERIAL CAME OUT OF THE CANNULA ONLY FOR ONE SYRINGE. A SMALL PORTION OF THIS MATERIAL WAS REMOVED FROM THE SAMPLE AND PREPARED FOR FTIR SPECTRAL ANALYSIS. THE SPECTRAL ANALYSIS SHOWS THAT THIS MATERIAL IS MOST LIKELY SILICONE. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 2213837. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. BASED ON THE SAMPLE RECEIVED, EMBECTA WAS ABLE TO CONFIRM THE CUSTOMER INDICATED ISSUE. A DEFINITIVE ROOT-CAUSE COULD NOT BE DETERMINED.
DATE OF EVENT IS UNKNOWN; AWARENESS DATE HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED WHILE USING RELION INSULIN SYRINGE 3/10ML 29G X 8MM FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PET OWNER REPORTED LIQUID COMING FROM SYRINGE WHEN DEPRESSING PLUNGER ROD, DURING PRIMING.
IT WAS REPORTED WHILE USING RELION INSULIN SYRINGE 3/10ML 29G X 8MM FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PET OWNER REPORTED LIQUID COMING FROM SYRINGE WHEN DEPRESSING PLUNGER ROD, DURING PRIMING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1407495 | RELION INSULIN SYRINGE 3/10ML 29G X 8MM | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 328512 | 2213837 | 00681131311786 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |