MENTOR MEMORYGEL BREAST IMPLANT
Report
- Report Number
- 1645337-2023-01878
- Event Type
- Injury
- Date Received
- February 21, 2023
- Date of Event
- September 20, 2021
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317000259
- PMA / PMN Number
- P030053
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ON MARCH 02, 2023, THE MENTOR ANALYSIS LAB RECEIVED THE SUSPECT MEDICAL DEVICE FOR EVALUATION. WITH THE RETURN OF THE DEVICE, THE LOT NUMBER WAS PROVIDED AS 6960908. ON MARCH 10, 2023, MENTOR COMPLETED AN EVALUATION ON THE RETURNED DEVICE. MENTOR CONDUCTED A VISUAL INSPECTION, MICROSCOPIC EXAMINATION, AND THICKNESS MEASUREMENT OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT THE BREAST IMPLANT WAS RUPTURED, RECEIVED IN TWO (2) PARTS, AND RECEIVED INCOMPLETE. MICROSCOPIC EXAMINATION WAS PERFORMED, AND THE CAUSE OF THE RUPTURE COULD NOT BE IDENTIFIED. THEREFORE, A THICKNESS MEASUREMENT WAS CONDUCTED ON THE SHELL AT THE TEAR, AND THE THICKNESS WAS WITHIN MANUFACTURING SPECIFICATIONS. ALTHOUGH NO CONCLUSION COULD BE REACHED ON THE CAUSE OF THE REPORTED EVENT, THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING CAUTION: RUPTURE CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. THE FOLLOWING THINGS MAY CAUSE THE IMPLANT TO RUPTURE: DAMAGE BY SURGICAL INSTRUMENTS; STRESSING THE IMPLANT DURING IMPLANTATION AND WEAKENING IT; FOLDING OR WRINKLING OF THE IMPLANT SHELL; EXCESSIVE FORCE TO THE CHEST (E.G. DURING CLOSED CAPSULOTOMY); TRAUMA; COMPRESSION DURING MAMMOGRAPHIC IMAGING; AND SEVERE CAPSULAR CONTRACTURE. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULE, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED ON LOT 6960908, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: CAPSULAR CONTRACTURE, RUPTURE MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A 51-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION SURGERY WITH IMPLANTATION OF A 225CC MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH BAKER GRADE III CAPSULAR CONTRACTURE OF THE RIGHT BREAST AND A RUPTURED RIGHT BREAST IMPLANT BY A MEDICAL PROFESSIONAL USING THE PATIENT'S SYMPTOMS. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT WITH 300CC MENTOR MEMORYGEL BREAST IMPLANTS ON (B)(6) 2022. THE DATE OF IMPLANTATION WAS PROVIDED TO MENTOR AS A BEST ESTIMATE. SEVERAL FOLLOW-UPS HAVE BEEN CONDUCTED, AND NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. IF MORE INFORMATION BECOMES AVAILABLE, MENTOR WILL SUBMIT A SUPPLEMENTAL REPORT ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 944031 | MENTOR MEMORYGEL BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 3502251BC | 6960908 | 00081317000259 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Female | Required Intervention |