FDA Adverse Event Injury Summary report: N

INCLUSIVE TAPERED IMPLANT 4.2 MMD X 8 MML X 3.5 MMP

MDR report key: 16406259 · Received February 20, 2023

Report

Report Number
3011649314-2023-00065
Event Type
Injury
Date Received
February 20, 2023
Date of Event
September 22, 2022
Report Date
January 29, 2025
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K121406
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CAPA 2023-006 THE DEVICE WAS RETURNED, THE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR LOT# 6071079 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: THERE WAS NO STOCK PRODUCT FROM LOT# 6071079 AVAILABLE FOR REVIEW. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE. THE IMPLANT WAS VERIFIED TO BE A INCLUSIVE TAPERED IMPLANT 4.2 MMD X 8 MML X 3.5 MMP (70-1070-IMP0026) USING RADIOGRAPHIC TEMPLATE (PK-209-062515). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. BONE DEBRIS WAS OBSERVED IN THE THREADING OF THE IMPLANT. THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART(S) BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE: "FAILURE TO OSSEOINTEGRATE" IS A COMMON COMPLAINT IN REGARDS TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR FAILURE TO OSSEOINTEGRATE IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE THE LOSS OF PRIMARY STABILITY AT THE OSTEOTOMY SITE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. IFU-3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "INCLUSIVE DENTAL IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE." IFU-3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU-3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." IFU-3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL AND COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU-3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN STERILITY SECTION: "INCLUSIVE DENTAL IMPLANTS ARE SHIPPED STERILE. THEY SHOULD NOT BE RESTERILIZED. THEY ARE FOR SINGLE USE ONLY, PRIOR TO EXPIRATION DATE."

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT BEEN RETURNED. IF/ WHEN THE DEVICE IS RETURNED AN INVESTIGATION WILL BE CARRIED OUT AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THE OTHER RELATED EVENT ARE CAPTURED IN THE FOLLOWING MEDWATCH REPORTS: 3011649314-2023-00064. 3011649314-2023-00066. 3011649314-2023-00067. CAPA CA-00016. MANUFACTURER REFERENCE: (B)(4).

Description of Event or Problem · 0

RECEIVED A PIQ ALONG WITH THE RMA ON (B)(6) 2023 FROM REGARDING A NEW COMPLAINT. IT WAS REPORTED THAT THE INCLUSIVE TAPERED IMPLANT FAILED. THE BONE GRADE IS NOTED AS GRADE III AND ORAL HYGIENE IS "FAIR." THE PATIENT HAS NO MEDICAL OR DENTAL HISTORY PRIOR TO IMPLANT. THE PATIENT PRESENTED ON 11-15-21 FOR IMPLANT PLACEMENT ON TOOTH #27. THE PATIENT RETURNED ON (B)(6) 2022 WITH COMPLAINTS OF PAIN. UPON EXAM THE PROVIDER NOTED INFLAMMATION. IT WAS DETERMINED THAT THE IMPLANT HAD LOSS INTEGRATION AND THE DEVICE WAS REMOVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INCLUSIVE TAPERED IMPLANT FAILED. THE BONE GRADE IS NOTED AS GRADE III AND ORAL HYGIENE IS FAIR. THE PATIENT HAS NO REPORTED MEDICAL OR DENTAL HISTORY PRIOR TO IMPLANT. THE PATIENT PRESENTED ON (B)(6) 2021 FOR IMPLANT PLACEMENT. THE SPECIFIC TOOTH FOR THE REPORTED LOT WAS NOT INDICATED BUT TEETH 19, 22, 27 AND 30 WERE MENTIONED. THE PATIENT RETURNED ON (B)(6) 2022, AFTER THE FINAL PROSTHESIS DELIVERY, WITH COMPLAINTS OF PAIN. UPON EXAMINATION, THE PROVIDER NOTED INFLAMMATION. IT WAS DETERMINED THAT THE IMPLANT HAD LOST INTEGRATION AND THE DEVICE WAS REMOVED AND NOT REPLACED. THE PATIENT'S SYMPTOMS RESOLVED AFTER IMPLANT REMOVAL. THERE WAS NO PERMANENT PATIENT INJURY AND NO ADDITIONAL MEDICAL/SURGICAL PROCEDURE WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1317839 INCLUSIVE TAPERED IMPLANT 4.2 MMD X 8 MML X 3.5 MMP INCLUSIVE TAPERED IMPLANT DZE PRISMATIK DENTALCRAFT, INC. 70-1070-IMP0026 6071079

Patients

Seq Age Sex Outcome Treatment
1 69 YR Female Required Intervention