INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2010-00134
- Event Type
- Malfunction
- Date Received
- March 18, 2010
- Date of Event
- February 19, 2010
- Report Date
- February 19, 2010
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- BZE
- PMA / PMN Number
- K034026
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). THE RETURNED BREATHING CIRCUIT WAS VISUALLY INSPECTED FOR SIGNS OF DAMAGE, SUCH AS HOLES OR TEARS. RESULTS: A 10MM LONG HOLE WAS FOUND IN THE EXPIRATORY EVAQUA BREATHING CIRCUIT LIMB. THE EDGE OF THE HOLE IS LACERATED. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT # 090612. CONCLUSION: BASED ON THE TYPE OF DAMAGE, THE CIRCUIT WAS PUNCTURED OR SCRATCHED BY A BLUNT OBJECT. THE HOLE CAUSED THE BREATHING CIRCUIT TO FAIL THE LEAK TEST. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS PRIOR TO DISTRIBUTION. ANY BREATHING CIRCUIT WHICH DOES NOT PASS THE PRESSURE TEST IS DISCARDED. THIS SUGGESTS THAT THE BREATHING CIRCUIT WAS PUNCTURED POST-PRODUCTION, POSSIBLY DURING STORAGE OR SET-UP. THE KEY DIFFERENCE BETWEEN FPH'S EVAQUA BREATHING CIRCUITS AND THE CONVENTIONAL BREATHING CIRCUITS IS THAT THE EXPIRATORY TUBE OF EVAQUA CIRCUITS SUCH AS THE RT340 IS COMPOSED OF A THIN, SEMI-PERMEABLE FILM SPECIALLY DESIGNED TO ALLOW WATER VAPOR FROM EXPIRED VENTILATORY GASES TO PASS THROUGH. THE EVAQUA EXPIRATORY TUBE HAS A PROTECTIVE MESH WHICH PREVENTS DAMAGE TO THE WALLS OF THE TUBE, HOWEVER, THE EVAQUA TUBING REMAINS MORE SUSCEPTIBLE TO DAMAGE THAN CONVENTIONAL CIRCUITS WHEN EXPOSED TO ROUGH HANDLING OR DAMAGE CAUSED BY SHARP OBJECTS AND CIRCUIT HANGERS. OUR USER INSTRUCTIONS ADVISE THE USER TO "FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE." WE HAVE RECEIVED TWO OTHER COMPLAINTS OF THIS TYPE IN THE PAST YEAR.
A HOSP IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE REP THAT A RT236 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS LEAKING AT THE Y-PIECE. THE REPORTED BREATHING CIRCUIT LEAK WAS FOUND PRIOR TO PT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT | BZE | FISHER & PAYKEL HEALTHCARE, LTD. | RT236 | 090612 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | MAQUET SERVO-I VENTILATOR |