UNKNOWN LOCKING SCREW
Report
- Report Number
- 0009610622-2023-00048
- Event Type
- Injury
- Date Received
- February 17, 2023
- Date of Event
- May 11, 2022
- Report Date
- May 4, 2023
- Manufacturer
- STRYKER TRAUMA KIEL
- Product Code
- HSB
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE REPORTED EVENT COULD BE CONFIRMED, FROM THE PROVIDED X-RAYS FOR EVALUATION AND MATCHES THE ALLEGED FAILURE MODE. THE DEVICE INSPECTION WAS NOT POSSIBLE AS THE PRODUCT WAS NOT RETURNED FOR INVESTIGATION. THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED BECAUSE THE AFFECTED DEVICE WAS NOT RETURNED, AND THE LOT NUMBER WAS NOT COMMUNICATED. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. A FEW X-RAYS WERE PROVIDED WHICH WERE PRESENTED TO OUR HEALTH CARE PROFESSIONAL FOR EVALUATION, AND A QUESTION ABOUT THE MOST LIKELY REASON FOR THIS EVENT WAS ASKED, TO WHICH THE MEDICAL EXPERT REPLIED: ¿THERE IS VERY LIMITED INFORMATION FOR ME TO WORK WITH HERE. NON-UNION IS CONFIRMED FOR THIS CASE. AS YOU MAY BE AWARE, NON-UNION IS OFTEN MULTI-FACTORIAL. ONE OF THE FACTORS IN THIS CASE MAY HAVE BEEN INSTABILITY OF THE CONSTRUCT FOR THIS LOW FEMUR FRACTURE. THE CONTINUOUS MOVEMENT IN THE NAIL, MIGHT HAVE CAUSED THE NON-UNION AND BREAKAGE OF THE SCREW.¿. [ORIGINAL STATEMENT(S). MORE DETAILED INFORMATION ABOUT THE COMPLAINT EVENT AS WELL AS THE AFFECTED DEVICE MUST BE AVAILABLE IN ORDER TO DETERMINE THE ROOT CAUSE OF THE COMPLAINT EVENT. IF THE DEVICE IS RETURNED OR ANY FURTHER INFORMATION IS PROVIDED, THE INVESTIGATION REPORT WILL BE REASSESSED.
BASED ON THE AVAILABLE INFORMATION THE DEVICE WILL NOT BE RETURNED THEREFORE AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED. A REVIEW OF THE DEVICE HISTORY IS NOT POSSIBLE BECAUSE THE LOT NUMBER WAS NOT COMMUNICATED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. DEVICE DISPOSITION UNKNOWN.
THE MANUFACTURER BECAME AWARE OF A TWITTER POST WHICH WAS PUBLISHED BY THE DR. Q IN MAY 2022. THE REPORT IS ABOUT A FEMORAL SHAFT FRACTURE THAT ENDED IN NON-UNION, WHICH IS ASSOCIATED WITH A STRYKER GAMMA 3 NAIL. THE POST CAN BE FOUND AT HTTPS://TWITTER.COM/AQUEIPOT/STATUS/1524466992899629056. IT WAS REPORTED THAT THERE WAS A NON-UNION 12 MONTHS AFTER IMPLANTATION. THE PATIENT HAD REPORTED PAIN, SO X-RAYS WERE TAKEN. ADDITIONALLY 2 SCREWS HAD BEEN BROKEN. THE NAIL AND SCREWS WERE REVISED AND THE NON-UNION HEALED AFTER 3 MONTHS.
THE MANUFACTURER BECAME AWARE OF A TWITTER POST WHICH WAS PUBLISHED BY THE DR. (B)(6) IN MAY 2022. THE REPORT IS ABOUT A FEMORAL SHAFT FRACTURE THAT ENDED IN NON-UNION, WHICH IS ASSOCIATED WITH A STRYKER GAMMA 3 NAIL. THE POST CAN BE FOUND AT HTTPS://TWITTER.COM/AQUEIPOT/STATUS/1524466992899629056. IT WAS REPORTED THAT THERE WAS A NON-UNION 12 MONTHS AFTER IMPLANTATION. THE PATIENT HAD REPORTED PAIN, SO X-RAYS WERE TAKEN. ADDITIONALLY 2 SCREWS HAD BEEN BROKEN. THE NAIL AND SCREWS WERE REVISED AND THE NON-UNION HEALED AFTER 3 MONTHS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 556435 | UNKNOWN LOCKING SCREW | ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES | HSB | STRYKER TRAUMA KIEL | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |