FDA Adverse Event Injury Summary report: N

LENSX LASER SYSTEM

MDR report key: 16397458 · Received February 17, 2023

Report

Report Number
2028159-2023-00216
Event Type
Injury
Date Received
February 17, 2023
Date of Event
January 25, 2023
Report Date
April 26, 2023
Manufacturer
ALCON LENSX, INC.
Product Code
OOE
UDI-DI
00380659981623
PMA / PMN Number
K163551
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. ALCON LENSX (SITE # (B)(4) IS NO LONGER OPERATIONAL. LENSX MANUFACTURED PRODUCTS ARE MAINTAINED AND INVESTIGATED BY THE ALCON RESEARCH, LTD. IRVINE TECHNOLOGY CENTER SITE # (B)(4). THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION PROVIDED IN H.3., H.6., AND H.10. A MANUFACTURING DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. SERVICE HISTORY WAS REVIEWED FOR THE SYSTEM. NO SERVICE RECORD RELEVANT TO THE COMPLAINT REPORTED EVENT WAS FOUND. HOWEVER, THE SYSTEM WAS LAST SERVICED PRIOR TO THE REPORTED EVENT PER SERVICE RECORD (SR). THE SYSTEM FOUND TO MEET ALL COSMETIC AND PERFORMANCE STANDARDS. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. THE CUSTOMER REPORTED EVENT CANNOT BE CONFIRMED. THUS, BASED ON THE INFORMATION OBTAINED, THE ROOT CAUSE OF THE REPORTED EVENT IS INCONCLUSIVE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A CUSTOMER REPORTED EXPERIENCED A POSTERIOR CAPSULE TEAR AND PERFORM A VITRECTOMY IN RIGHT EYE OF A PATIENT, DURING SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
162599 LENSX LASER SYSTEM OPHTHALMIC FEMTOSECOND LASER OOE ALCON LENSX, INC. 550 NA 00380659981623

Patients

Seq Age Sex Outcome Treatment
1 Male Other