FDA Adverse Event Injury Summary report: N

BODYTITE

MDR report key: 16395050 · Received February 17, 2023

Report

Report Number
3010511300-2023-00067
Event Type
Injury
Date Received
February 17, 2023
Date of Event
December 2, 2021
Report Date
February 17, 2023
Manufacturer
INMODE LTD
Product Code
GEI
UDI-DI
17290016633211
PMA / PMN Number
K182325
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

USER ERROR: EXCESSIVE QUANTITY OF ENERGY WAS DELIVERED (25.2KJ) IN RELATIVELY SMALL AREA (SUPRAPUBIC AREA- MONS). THIS ENERGY SURPLUS, TRANSFORMING INTO HEAT, WAS IRRADIATED WITH A RETARDING EFFECT TOWARDS THE DERMAL TISSUE, CAUSING BURNS AND TISSUE NECROSIS. A SUBSEQUENT SECONDARY INFECTION CANNOT BE RULED OUT. SUMMARY OF RELATED INVESTIGATIONS AN EXCESSIVE AMOUNT OF ENERGY GENERATED AN ACCUMULATED SURPLUS OF HEAT WHICH EVOLVED INTO TRANS SKIN BURN WITH NECROTIC OUTCOME AND POSSIBLY A SECONDARY INFECTION (ERYSIPELAS).

Additional Manufacturer Narrative · 0

FOLLOW UP REPORT 07/05/2023. 1. H6. HEALTH EFFECT - IMPACT CODE - CHANGED TO 4619. MEDICAL DEVICE PROBLEM CODE - CHANGED TO 1126. INVESTIGATION FINDINGSCODE - CHANGED TO 4248. INVESTIGATION CONCLUSION CODE - CHANGED TO 61. 2. H11. ATTACHED: UPON ARRIVAL TEST. FINAL TEST. IN SERVICE REPORT (RETRAIN).

Description of Event or Problem · 0

A TISSUE NECROSIS IN THE SUPRAPUBIC REGION (MONS) FOLLOWING A REDUCTION TREATMENT WITH BODYTITE. SUSPICIOUS FOR ERYSIPELAS AS THE UNDERLYING TISSUE BLANCHED FEW DAYS AFTER. A WEEK LATER, THE WOUNDS EVOLVED AND PSEUDOESCHAR WAS SEEN. IN THE LAST PHOTO THERE IS SIGNIFICANT STRAW-COLORED DRAINAGE, POSSIBLY A SMALL SEROMA DRAINING FROM A FULL THICKNESS WOUND.

Description of Event or Problem · 0

A TISSUE NECROSIS IN THE SUPRAPUBIC REGION (MONS) FOLLOWING A REDUCTION TREATMENT WITH BODYTITE. SUSPICIOUS FOR ERYSIPELAS AS THE UNDERLYING TISSUE BLANCHED FEW DAYS AFTER. A WEEK LATER, THE WOUNDS EVOLVED AND PSEUDOESCHAR WAS SEEN. IN THE LAST PHOTO THERE IS SIGNIFICANT STRAW-COLORED DRAINAGE, POSSIBLY A SMALL SEROMA DRAINING FROM A FULL THICKNESS WOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
170569 BODYTITE FACETITE GEI INMODE LTD AG606972A 17290016633211

Patients

Seq Age Sex Outcome Treatment
1 32 YR Female Required Intervention