ADULT INSPIRATORY HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2010-00150
- Event Type
- Other
- Date Received
- March 18, 2010
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- BTT
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
(B)(4). METHOD: THE DISTRIBUTOR REPORTED THAT A BOX OF RT114 BREATHING CIRCUITS WAS LABELED WITH THE INCORRECT PRODUCT LABEL (RT203). THE PRODUCT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE'S REGIONAL OFFICE IN (B)(4) WHO HANDLED THE PRODUCT DISPOSITION PER THE RELEVANT COMPANY PROCEDURES. AN IMAGE OF THE LABEL ON THE CIRCUIT BOX WAS PROVIDED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR ANALYSIS. RESULTS: THE IMAGE PROVIDED SHOWS AN RT203 PRODUCT LABEL. FISHER & PAYKEL HEALTHCARE'S REGIONAL OFFICE IN (B)(4) CONFIRMED THAT THE PACKAGE CONTAINED RT114 BREATHING CIRCUITS WITH LOT NUMBER 100112. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 100112. CONCLUSION: THE PACKAGING HAD BEEN PRE-LABELED WITH AN RT203 PRODUCT LABEL, HOWEVER, THE MFG LINE CHANGED OVER TO THE RT114 BREATHING CIRCUIT. THIS RESULTED IN THE RT114 BREATHING CIRCUIT KITS BEING PACKED IN A BOX LABELED RT203. THERE ARE STANDARD OPERATING PROCEDURES (SOPS) IN PLACE TO ASSIST OPERATORS ON THE PRODUCTION LINE CORRECTLY PACK BREATHING CIRCUITS. THIS INCLUDES INSTRUCTIONS STATING THAT A LABEL SHOULD ONLY BE AFFIXED TO A BOX IF THE BOX IS GOING TO BE PACKED. (B)(4).
A DISTRIBUTOR IN (B)(4) REPORTED THAT THEY RECEIVED A BOX OF RT114 ADULT INSPIRATORY HEATED BREATHING CIRCUITS WITH AN RT203 PRODUCT LABEL. THE MISLABELED PACKAGE WAS DISCOVERED BY THE DISTRIBUTOR PRIOR TO CUSTOMER DISTRIBUTION OR PT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT INSPIRATORY HEATED BREATHING CIRCUIT | BTT | FISHER & PAYKEL HEALTHCARE, LTD. | RT114 | 100112 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |