FDA Adverse Event
No answer provided
Summary report: N
AVIEW 2 ADVANCE
MDR report key: 16389373
·
Received February 16, 2023
Report
- Report Number
- 1220828-2023-00004
- Event Type
- No answer provided
- Date Received
- February 16, 2023
- Report Date
- February 17, 2023
- Manufacturer
- AMBU A/S
- Product Code
- FEM
- Report Source
- Distributor report
- Reporter Location
- AS
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
THE MONITOR WAS NOT RECEIVED FOR INVESTIGATION SINCE IT WAS DISPOSED OF BY FIRE BRIGADE. PICTURES WERE PROVIDED. INVESTIGATION IS STILL ONGOING. THE EVENT IS SUSPECTED TO BE CAUSED BY THE IMPACT ON THE BATTERY WHEN THE MONITOR WAS DROPPED ONTO THE FLOOR.
Description of Event or Problem · 0
THE AVIEW 2 ADVANCE DISPLAY UNIT WAS DROPPED FROM APPROXIMATELY 1.5 METERS. SPARKS WAS OBSERVED AND THE MONITOR CAUGHT FIRE. THE DISPLAY LEFT A BURN MARK ON THE FLOOR. FIRE BRIGADE WAS CALLED TO HANDLE THE MONITOR AND DISPOSED IT. PATIENT WAS NOT INVOLVED. NO HARM TO USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 358369 | AVIEW 2 ADVANCE | AVIEW 2 ADVANCE | FEM | AMBU A/S | 405011000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |