FDA Adverse Event Malfunction Summary report: N

COBAS C 503 ANALYTICAL UNIT

MDR report key: 16388841 · Received February 16, 2023

Report

Report Number
1823260-2023-00481
Event Type
Malfunction
Date Received
February 16, 2023
Date of Event
January 19, 2023
Report Date
April 3, 2023
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
UDI-DI
07613336121528
PMA / PMN Number
K191899
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

REPORTEDLY THE CALIBRATION AND QUALITY CONTROLS WERE ACCEPTABLE AT THE TIME OF TESTING. THE CUSTOMER WAS CONCERNED THE RESULT HAD A "HU" DATA FLAG. PER PRODUCT LABELING: "HU - ALARM: HIGHER UNCERTAINTY, DESCRIPTION: THE RESULT LIES BETWEEN THE LIMIT OF DETECTION (LOD) AND THE LIMIT OF QUANTITATION (LOQ)., CAUSE: THE SAMPLE CONCENTRATION IS TOO LOW., REMEDY: 1. FOR SPECIFIC C 503 TESTS: REPEAT THE TEST WITH INCREASED SAMPLE VOLUME, IF AVAILABLE FOR THE TEST. CHECK THE SAMPLE VOLUMES ON MENU > APPLICATION. 2. FOR ISE AND C 503 TESTS: PERFORM MAINTENANCE > 18 SAMPLE PROBE WASH. 3. REPEAT THE TEST WITH A NEW ALIQUOT." THE CUSTOMER DID NOT PROVIDE ANY ADDITIONAL INFORMATION FOR THE INVESTIGATION. THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Description of Event or Problem · 0

THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE TP2 (TOTAL PROTEIN GEN. 2) ASSAY ON A COBAS C 503 ANALYTICAL UNIT. THE SAMPLE INITIALLY RESULTED IN A TP2 VALUE OF 0.3 G/DL, ACCOMPANIED BY A DATA FLAG. THE SAMPLE WAS REPORTED OUTSIDE OF THE LABORATORY AND QUESTIONED. THE SAMPLE WAS REPEATED RESULTING IN A VALUE OF 7.0 G/DL, WHICH WAS DEEMED CORRECT. THE TP2 REAGENT LOT WAS 63565201 WITH AN EXPIRATION DATE OF 31-JUL-2023.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
373750 COBAS C 503 ANALYTICAL UNIT CLINICAL CHEMISTRY ANALYZER JJE ROCHE DIAGNOSTICS C503 NA 07613336121528

Patients

Seq Age Sex Outcome Treatment
1 Unknown