FDA Adverse Event Malfunction Summary report: N

ENDOPATH ETS

MDR report key: 163870 · Received April 20, 1998

Report

Report Number
1527736-1998-01216
Event Type
Malfunction
Date Received
April 20, 1998
Date of Event
March 23, 1998
Report Date
March 24, 1998
Manufacturer
ETHICON ENDO-SURGERY
Product Code
KOG
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TRACKING #.50744. EES #.982090/C. H6; CODE 400: BENT CARTRIDGE LOCKOUT TAB. ENDOPATH ETS: BASED UPON THE INQUIRY INFO RECEIVED, THE VISUAL EXAMINATION, AND THE FUNCTIONAL TESTING, NO CONCLUSION COULD BE REACHED AS TO WHY THE INSTRUMENT REPORTEDLY, "COULD NOT BE FIRED" DURING SURGERY. THE INSTRUMENT WAS RETURNED IN GOOD PHYSICAL CONDITION. THE INSTRUMENT WAS CYCLED, FIRED, CUT, AND FORMED THE STAPLES WITHIN DESIGN SPECIFICATION. IT WAS CONCLUDED THAT THE INSTRUMENT WAS FULLY FUNCTIONAL AND CONFORMING TO DESIGN SPECIFICATIONS. THE EXPERIENCE THE SURGEON REPORTED COULD NOT BE REPEATED. THE RETURNED CARTRIDGE HAD A BENT LOCKOUT TAB ON THE PAN WHICH INDICATES THAT THE INSTRUMENT'S FIRING CYCLE WAS INTERRUPTED, RELEASED, THEN RESTARTED. WHEN THIS OCCURRED, THE LOCKOUT TAB ON THE CARTRIDGE BECAME DAMAGED.

Description of Event or Problem · 1

IT WAS REPORTED BY THE REP THE DEVICE WAS USED DURING A LAPAROSCOPIC ASSISTED VAGINAL HYSTERECTOMY. IT WAS REPORTED THE TSW35 WOULD NOT FIRE. NO OTHER INFO WAS AVAILABLE. THERE WAS NO CONSEQUENCE TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS LINEAR CUTTER KOG ETHICON ENDO-SURGERY NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other