PREPSTAIN SYSTEM
Report
- Report Number
- 1062336-2010-00004
- Event Type
- Malfunction
- Date Received
- March 16, 2010
- Date of Event
- February 17, 2010
- Manufacturer
- ANDREAS HETTICH GMBH & CO.
- Product Code
- MKQ
- PMA / PMN Number
- P970018
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ON MARCH 4, 2010, (B)(6), NOTIFIED BD TRIPATH OF A FIELD ACTION REGARDING THE ROTINA 380 MODELS. ON THE LID OF THE CENTRIFUGE THERE ARE TWO "GAS SPRINGS" THAT ASSIST THE LID TO BE RAISED AND PREVENT IT FROM FALLING WHEN RELEASED. THE SPRINGS ARE ATTACHED TO THE LID AT ONE END AND TO A BRACKET INSIDE THE CENTRIFUGE HOUSING AT THE OTHER. THE CENTRIFUGE WAS RECEIVED ON MARCH 9, 2010. QA ENGINEERING AND OPERATIONS EVALUATED THE RETURNED CENTRIFUGE AND CONFIRMED THE FIELD ACTION (RECALL/RETROFIT) DEFECT CITED BY (B)(6). BOTH GAS STRUTS WERE DETACHED, HOWEVER, THE LEFT REAR STRUT APPARENTLY HAD TOUCHED THE POWER TRANSFORMER CAUSING AN ELECTRICAL SHORTAGE, BECAUSE ELECTRICAL BURNS COULD BEEN SEEN ON THE STRUT AND TRANSFORMER. ALSO, THE INSTRUMENT WOULD NOT POWER-UP, THEREFORE, NO OTHER PARAMETERS COULD BE EVALUATED. (B)(6) HAS INITIATED AN INVESTIGATION TO DETERMINE THE ROOT CAUSE AND APPROPRIATE CORRECTIVE ACTIONS. IN ADDITION, (B)(6) IS FACILITATING ALL ACTIVITIES ASSOCIATED WITH THEIR FIELD ACTION. BD DIAGNOSTICS HAS PROVIDED THE LOCATIONS AND CUSTOMERS FOR THE AFFECTED INSTRUMENTS.
THE CYTOPATHOLOGY LABORATORY OBSERVED THAT ONE OF THE SHOCKS WAS DAMAGED ON THE CENTRIFUGE. THE LID WILL NOT STAY UP. IN ADDITION, THERE WAS A BURNING SMELL COMING FROM THE CENTRIFUGE SO THEY DISCONNECTED THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PREPSTAIN SYSTEM | HETTICH CENTRIFUGE (COMPONENT) | MKQ | ANDREAS HETTICH GMBH & CO. | ROTINA 380 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |