LP10 VOLUME VENTILATOR
MDR report key: 163805
·
Received April 17, 1998
Report
- Report Number
- 2183157-1998-00065
- Event Type
- Malfunction
- Date Received
- April 17, 1998
- Date of Event
- March 26, 1998
- Report Date
- April 17, 1998
- Manufacturer
- NELLCOR PURITAN BENNETT
- Product Code
- CBK
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
TESTING BY MFR FOUND UNIT TO NOT CYCLE WITH ALL LEDS AND CONSTANT SINGLE TONE ALARM DUE TO INTEGRATED CIRCUIT U28 ON THE LOGIC BOARD BEING OUT OF SPECIFICATION. REPLACED U28.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LP10 VOLUME VENTILATOR | VOLUME VENTILATOR | CBK | NELLCOR PURITAN BENNETT | LP10 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |