FDA Adverse Event Malfunction Summary report: N

LP10 VOLUME VENTILATOR

MDR report key: 163805 · Received April 17, 1998

Report

Report Number
2183157-1998-00065
Event Type
Malfunction
Date Received
April 17, 1998
Date of Event
March 26, 1998
Report Date
April 17, 1998
Manufacturer
NELLCOR PURITAN BENNETT
Product Code
CBK
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

TESTING BY MFR FOUND UNIT TO NOT CYCLE WITH ALL LEDS AND CONSTANT SINGLE TONE ALARM DUE TO INTEGRATED CIRCUIT U28 ON THE LOGIC BOARD BEING OUT OF SPECIFICATION. REPLACED U28.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LP10 VOLUME VENTILATOR VOLUME VENTILATOR CBK NELLCOR PURITAN BENNETT LP10 NA

Patients

Seq Age Sex Outcome Treatment
1 * Other