INTELLANAV STABLEPOINT OPEN-IRRIGATED
Report
- Report Number
- 2124215-2023-05540
- Event Type
- Malfunction
- Date Received
- February 15, 2023
- Date of Event
- January 18, 2023
- Report Date
- February 15, 2023
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- OAE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
INTELLANAV STABLEPOINT OPEN-IRRIGATED CATHETER WAS EVALUATED BY BOSTON SCIENTIFIC CORPORATION. VISUAL INSPECTION NOTED THAT THE CATHETER SHAFT WAS DAMAGED AT DISTAL SECTION. A FUNCTIONAL TEST WAS PERFORMED AND THE STEERING KNOB AND THE TENSION CONTROL KNOB FUNCTIONED PROPERLY ON BOTH LOCK AND UNLOCK POSITIONS. NO ABNORMAL RESISTANCE WAS FELT WHEN ACTUATING THE STEERING MECHANISM. LEAKAGE TEST WAS DONE; THE LUMEN OF THE DEVICE WAS LEAK TESTED USING AN ISAAC HD LEAK TESTER (PRESSURE DECAY TEST SYSTEM) AND A TOUHY BOURST SEAL FOR SEALING THE IRRIGATION HOLES AND MINI ELECTRODES. THE LUMEN PRESSURE DECAY WAS MEASURED THREE TIMES. THE UNIT PASSED THE TEST WITHOUT PROBLEMS. FINALLY, A FLOW TEST WAS CONDUCTED; THE DEVICE WAS CONNECTED TO A METRIQ PUMP AND WAS PURGED AT 60 ML/MIN. ALL SIX IRRIGATION PORTS FLOW FREELY. THE PUMP WAS PROGRAMMED TO 30ML/MIN AND THE DEVICE WAS IRRIGATED FOR 5 MINUTES WITHOUT ANY ERRORS OR PUMP MESSAGES. THE DEVICE WAS FOUND WITHIN SPECIFICATIONS. LABORATORY ANALYSIS WAS UNABLE TO CONFIRM THE REPORTED CLINICAL OBSERVATIONS.
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE TO TREAT AN ATRIAL TACHYCARDIA, AN INTELLANAV STABLEPOINT OPEN-IRRIGATED CATHETER WAS SELECTED FOR USE. FLOW RATE WAS 60ML/MIN. AN INSUFFICIENT IRRIGATION AMOUNT WHEN FLUSHING WAS PERFORMED AS USUAL. CATHETER WAS REPLACED AND THE ISSUE WAS RESOLVED. PROCEDURE WAS ABLE TO BE COMPLETED SUCCESSFULLY WITHOUT ANY PATIENT COMPLICATIONS. CATHETER WAS RETURNED TO BOSTON SCIENTIFIC CORPORATION FOR LABORATORY ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1290225 | INTELLANAV STABLEPOINT OPEN-IRRIGATED | CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION | OAE | BOSTON SCIENTIFIC CORPORATION | 0029560820 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |