FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 16379313 · Received February 15, 2023

Report

Report Number
1314492-2023-00408
Event Type
Malfunction
Date Received
February 15, 2023
Report Date
March 23, 2023
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
UDI-DI
00085412498683
PMA / PMN Number
K133801
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

CORRECTED INFORMATION H6: DEVICE PROBLEM CODE A160105 IS CORRECTED TO A160106. ADDITIONAL INFORMATION H3, H6 AND H10: A SERVICE HISTORY REVIEW REVEALED NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE DEVICE WAS NOT RECEIVED FOR EVALUATION; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SPECTRUM PUMP FAILED A DOWNSTREAM OCCLUSION POUNDS PER SQUARE INCH TEST IN THE BIOMEDICAL SERVICE DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1126803 SPECTRUM INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE CORPORATION NA NA 00085412498683

Patients

Seq Age Sex Outcome Treatment
1 Unknown