FDA Adverse Event Injury Summary report: N

SNAPFIX SCREW FULL THREADED Ø2.0MM - LENGTH 12MM

MDR report key: 16378842 · Received February 15, 2023

Report

Report Number
3012966183-2023-00003
Event Type
Injury
Date Received
February 15, 2023
Date of Event
January 13, 2023
Report Date
February 15, 2023
Manufacturer
TYBER MEDICAL
Product Code
HWC
PMA / PMN Number
K153575
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

DEVICE NOT RETURNED; THEREFORE, A DEVICE EVALUATION WAS NOT CONDUCTED. PHOTOGRAPHIC EVIDENCE WAS PROVIDED WHICH SHOWED THE DEVICE WAS FRACTURED. DEVICE HISTORY RECORD REVIEW WAS CONDUCTED . LOT 226421 WAS MFG. 8-5-2021. LOT 226421 HAD 250 PIECES. THE LOT DID NOT HAVE ANY NON-CONFORMANCES OR CAPAS. LOT MET ALL SPECIFICATIONS. HISTORICAL DATA REVIEW IDENTIFIED 1 OTHER COMPLAINT RELATED TO LOT 226421 FOR A SIMILAR ISSUE. THERE IS NO IMAGING, X-RAYS, FLUOROSCOPY, MRI ETC. OR ADDITIONAL INFORMATION PROVIDED BY THE COMPLAINANT THAT CAN CONFIRM THE COMPLAINT EVENT. THE ROOT CAUSE IS INDETERMINATE.

Description of Event or Problem · 0

COULDN'T SNAP OFF THE STEM ON THE SNAP OFF WHICH RESULTED IN FRACTURING THE OSTEOTOMY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1317562 SNAPFIX SCREW FULL THREADED Ø2.0MM - LENGTH 12MM SCREW HWC TYBER MEDICAL FSS212 226421

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other