PUREPOINT LASER
Report
- Report Number
- 2028159-2023-00196
- Event Type
- Malfunction
- Date Received
- February 15, 2023
- Date of Event
- January 17, 2023
- Report Date
- April 3, 2023
- Manufacturer
- ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
- Product Code
- HQF
- UDI-DI
- 00380657505975
- PMA / PMN Number
- K062624
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- 003
Narratives
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
THE COMPANY REPRESENTATIVE, WAS ABLE TO REPLICATE THE REPORTED EVENT. THE REPRESENTATIVE PERFORMED, POWER CALIBRATION AND ALIGNMENT. THEN THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. A NON-CONFORMANCE BASED REVIEW OF THE SERIAL NUMBER WAS PERFORMED AND DID NOT REVEAL, ANY POTENTIAL CONTRIBUTING FACTORS TO THE REPORTED COMPLAINT. A MANUFACTURING DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED, PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. A REVIEW FOR COMPLAINTS REPORTED AGAINST THIS SERIAL NUMBER WAS PERFORMED. NO SIMILAR COMPLAINTS WERE REPORTED FOR THE PRODUCT SERIAL UNDER INVESTIGATION. THE ROOT CAUSE OF THE REPORTED EVENT IS ATTRIBUTED TO THE POWER BEING OUT OF CALIBRATION. HOW OR WHY THIS OCCURRED, CANNOT BE CONCLUSIVELY DETERMINED. MANUFACTURER WILL CONTINUE TO MONITOR DATA FOR EVIDENCE OF ADVERSE TRENDING AND TAKE FURTHER ACTION, AS APPROPRIATE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
A CUSTOMER REPORTED THAT AN OPHTHALMIC OPERATING LASER WAS UNABLE TO MAKE THE MARK. PROCEDURE DETAILS AND PATIENT IMPACT WERE NOT REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1126714 | PUREPOINT LASER | LASER, OPHTHALMIC | HQF | ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER | NA | NA | 00380657505975 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |