FDA Adverse Event Injury Summary report: N

INION OTPS MINI PLATING SYSTEM

MDR report key: 1637744 · Received March 19, 2010

Report

Report Number
9710629-2010-00001
Event Type
Injury
Date Received
March 19, 2010
Date of Event
May 22, 2009
Report Date
March 18, 2010
Manufacturer
INION OY
Product Code
MAI
PMA / PMN Number
K023887
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IN THIS CASE THE CAST HAD BEEN REMOVED TOO EARLY, I.E., ONLY 3 WEEKS POSTOPERATIVELY (VS TYPICAL BONE HEALING TIME OF 6 WEEKS), AND THEREFORE, THE ROOT CAUSE OF THIS INCIDENT IS IMPROPER USE. NOTE THAT THE INSTRUCTIONS FOR USE OF INION OTPS MINI STATES THAT THE DEVICE IS USED IN CONJUNCTION WITH APPROPRIATE IMMOBILIZATION: INDICATIONS: THESE INION OTPS BIODEGRADABLE FIXATION SYSTEM IMPLANTS ARE INTENDED FOR USE IN FIXATION OF NON-COMMINUTED DIAPHYSEAL FRACTURES OF THE METACARPAL, PROXIMAL PHALANGEAL, MIDDLE PHALANGEAL AND OSTEOTOMIES IN THE PRESENCE OF APPROPRIATE IMMOBILIZATION. CONTRAINDICATIONS: THESE INION OTPS BIODEGRADABLE FIXATION SYSTEM IMPLANTS ARE NOT INTENDED FOR USE IN AND ARE CONTRAINDICATED FOR: ACTIVE OR POTENTIAL INFECTION. PATIENT CONDITIONS INCLUDING LIMITED BLOOD SUPPLY, INSUFFICIENT QUANTITY OR QUALITY OF BONE; AND WHERE PATIENT COOPERATION CANNOT BE GUARANTEED (E.G., ALCOHOLISM, DRUG ABUSE). LOAD BEARING PROCEDURES.

Description of Event or Problem · 1

ACCORDING TO THE RECEIVED INFORMATION, SURGEON PERFORMED AN OPERATION WITH INION OTPS MINI HAND PLATES USING INION OPTS 2.0 MM SCREWS ON (B) (6) 2009. A CAST WAS APPLIED. THE CAST WAS REMOVED ON (B) (6) 2009. DURING A VISIT ON (B) (6) 2009, IT WAS NOTED THAT THE FIXATION HAD FAILED, AND A REOPERATION WAS DONE WITH METAL IMPLANTS ON (B) (6) 2010. DURING REOPERATION, IT WAS NOTICED THAT THE BIODEGRADABLE PLATE WAS BENT AND THE SCREWS WERE BROKEN. THE PATIENT OUTCOME IS REPORTED GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INION OTPS MINI PLATING SYSTEM BONE PLATE, BONE SCREW, BIODEGRADABLE MAI INION OY UNK UNK

Patients

Seq Age Sex Outcome Treatment
1 41 YR Required Intervention