ADVANTA2 RENTAL BED
Report
- Report Number
- 3006697241-2023-00010
- Event Type
- Malfunction
- Date Received
- February 14, 2023
- Date of Event
- January 16, 2023
- Report Date
- February 14, 2023
- Manufacturer
- HILL-ROM MEXICO
- Product Code
- FNJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
THE HILLROM TECHNICIAN FOUND THE FRAME NEEDED TO BE CALIBRATE. PER THE HILLROM SERVICE MANUAL THE ADVANTA¿ 2 SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. TEST EACH OF THE BUTTONS TO CHECK THAT THEY ACTIVATE THE CORRECT FUNCTION AND THAT THEY DO NOT WORK INTERMITTENTLY BY PRESSING EACH BUTTON FOR SEVERAL SECONDS. EACH MOVEMENT MUST BE CONTINUOUS. REPLACE THE PENDANT IF NECESSARY. A SEARCH OF THE HILLROM MAINTENANCE RECORDS SHOWED HILLROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN DECEMBER 2022. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN CALIBRATED THE BED FRAME TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILLROM RECEIVED A REPORT FROM A HILLROM TECHNICIAN STATING THE HEAD OF BED WAS MOVING ON ITS OWN. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1516972 | ADVANTA2 RENTAL BED | BED, MANUAL | FNJ | HILL-ROM MEXICO | P1190ARENT01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |