FDA Adverse Event Malfunction Summary report: N

ADVANTA2 RENTAL BED

MDR report key: 16375287 · Received February 14, 2023

Report

Report Number
3006697241-2023-00010
Event Type
Malfunction
Date Received
February 14, 2023
Date of Event
January 16, 2023
Report Date
February 14, 2023
Manufacturer
HILL-ROM MEXICO
Product Code
FNJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE HILLROM TECHNICIAN FOUND THE FRAME NEEDED TO BE CALIBRATE. PER THE HILLROM SERVICE MANUAL THE ADVANTA¿ 2 SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. TEST EACH OF THE BUTTONS TO CHECK THAT THEY ACTIVATE THE CORRECT FUNCTION AND THAT THEY DO NOT WORK INTERMITTENTLY BY PRESSING EACH BUTTON FOR SEVERAL SECONDS. EACH MOVEMENT MUST BE CONTINUOUS. REPLACE THE PENDANT IF NECESSARY. A SEARCH OF THE HILLROM MAINTENANCE RECORDS SHOWED HILLROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN DECEMBER 2022. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN CALIBRATED THE BED FRAME TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 0

HILLROM RECEIVED A REPORT FROM A HILLROM TECHNICIAN STATING THE HEAD OF BED WAS MOVING ON ITS OWN. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1516972 ADVANTA2 RENTAL BED BED, MANUAL FNJ HILL-ROM MEXICO P1190ARENT01

Patients

Seq Age Sex Outcome Treatment
1 Unknown