FDA Adverse Event No answer provided Summary report: N

MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)

MDR report key: 16371638 · Received February 14, 2023

Report

Report Number
16371638
Event Type
No answer provided
Date Received
February 14, 2023
Date of Event
January 15, 2023
Report Date
January 25, 2023
Manufacturer
PHILIPS NORTH AMERICA LLC
Product Code
MHX
Report Source
User Facility report
Reporter Location
PA
Reporter Occupation
RISK MANAGER
Health Professional
N

Narratives

Description of Event or Problem · 0

WE EXPERIENCED A POWER OUTAGE WHICH LED TO A PHILLIPS MONITOR ISSUE. NURSING CAN VIEW RHYTHMS AND VITAL SIGNS FROM THE CENTRAL MONITOR, BUT MONITORS ARE NOT COMMUNICATING WITH EPIC. UPDATE FROM UF: WE FOUND OUT FROM PHILIPS THAT THE ISSUE WAS WITH A SERVICE/PROCESS ON THE SERVER STOPPING INTERMITTENTLY AND RESTARTING WHICH TAKES ABOUT 5 SECONDS TO REACH FUNCTIONAL STATE. THE SERVICE STOPPING ISSUE HAPPENS WHEN HIGH LOG ON/LOG/OFF ACTIVITY OCCURS VIA THE CARE ASSIST APP POSSIBLY AT SHIFT CHANGES. PHILIPS REPORTED THAT THEY HAD KNOWN ABOUT THIS ISSUE AS EARLY AS JULY 2022.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
648560 MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) MHX PHILIPS NORTH AMERICA LLC

Patients

Seq Age Sex Outcome Treatment
1 Unknown