MAGNUM INSTRUMENT
Report
- Report Number
- 2020394-2023-00064
- Event Type
- Malfunction
- Date Received
- February 14, 2023
- Date of Event
- January 18, 2023
- Report Date
- February 21, 2023
- Manufacturer
- BARD PERIPHERAL VASCULAR, INC.
- Product Code
- KNW
- UDI-DI
- 00801741083945
- PMA / PMN Number
- K934371
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
AS THE SERIAL NUMBER FOR THE DEVICE WAS PROVIDED, A REVIEW OF THE DEVICE HISTORY RECORDS IS CURRENTLY BEING PERFORMED. THE RETURN OF THE SAMPLE IS PENDING. THE INVESTIGATION OF THE REPORTED EVENT IS CURRENTLY UNDERWAY. (EXPIRY DATE: 12/2050).
H10: MANUFACTURING REVIEW: A DEVICE HISTORY RECORD AND MANUFACTURING REVIEW WAS NOT REQUIRED AS THE EVENT WAS DETERMINED TO BE EXPECTED AND THE INVESTIGATION DID NOT IDENTIFY ANY MANUFACTURING AND/OR SERVICE-RELATED ISSUES. INVESTIGATION SUMMARY: THE MAGNUM DRIVER WAS RETURNED FOR EVALUATION. THE MAGNUM DRIVER WAS VISUALLY INSPECTED UPON RECEIPT AND WAS FOUND TO BE IN GOOD CONDITION. ALSO DISCOLORED PLASTICS IDENTIFIED DURING EVALUATION. THE DEVICE WAS FUNCTIONALLY TESTED AND FAILED THE TESTS AS BENT SEQUENCER WELDMENT ASSEMBLY TAB. NO OTHER ANOMALIES WERE IDENTIFIED. THEREFORE, THE INVESTIGATION IS DETERMINED TO BE CONFIRMED FOR THE REPORTED DEVICE DOESN'T PRIME CORRECTLY. THE ROOT CAUSE FOR THE REPORTED DEVICE DOESN'T PRIME CORRECTLY WAS DETERMINED TO BE BENT SEQUENCER WELDMENT ASSEMBLY TAB. LABELING REVIEW: A REVIEW OF PRODUCT LABELING DOCUMENTATION (E.G., PROCEDURAL INSTRUCTIONS, INDICATIONS, WARNINGS, PRECAUTIONS, CAUTIONS, POSSIBLE COMPLICATIONS, CONTRAINDICATIONS, NURSING GUIDE, AND UNIT LABEL) DID NOT FIND ANY PRODUCT LABELING INADEQUACY. H10: D4 (EXPIRY DATE: 12/2050), G3. H11: H6 (METHOD, RESULT, CONCLUSION). H11: SECTION A THROUGH F: THE INFORMATION PROVIDE BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE DEVICE DOESN'T PRIME CORRECTLY. THERE WAS NO REPORTED PATIENT INJURY.
IT WAS REPORTED THAT DURING A BIOPSY PROCEDURE, THE DEVICE DOESN'T PRIME CORRECTLY. THERE WAS NO REPORTED PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2211436 | MAGNUM INSTRUMENT | BIOPSY INSTRUMENT | KNW | BARD PERIPHERAL VASCULAR, INC. | MG1522 | 00801741083945 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |