FDA Adverse Event Injury Summary report: N

PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM

MDR report key: 16363969 · Received February 13, 2023

Report

Report Number
2024168-2023-01436
Event Type
Injury
Date Received
February 13, 2023
Date of Event
January 20, 2023
Report Date
April 7, 2023
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
UDI-DI
08717648235184
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED ¿NO BLOOD OBSERVED AT THE MARKER LUMEN¿ WAS NOT CONFIRMED. A REVIEW OF THE MANUFACTURING RECORDS REVEALED NO MANUFACTURING NONCONFORMITIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. THE INVESTIGATION DETERMINED THE REPORTED DIFFICULTIES AND TREATMENT APPEAR TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.D4 - LOT # UPDATED FROM 2100241 TO: 2100741

Additional Manufacturer Narrative · 0

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS A VENOUS CLOSURE OF THE RIGHT COMMON FEMORAL VEIN USING THE PRE-CLOSE TECHNIQUE VIA AN 8F SHEATH HOLE PRIOR TO A MICRA PACEMAKER IMPLANTATION PROCEDURE. REPORTEDLY, THE PROSTYLE DEVICE WAS SUCCESSFULLY PRE-FLUSHED PRIOR TO USE. AFTER INSERTION, THERE WAS NO BLOOD OBSERVED AT THE MARKER LUMEN. THE SUTURES OF TWO NEW PROSTYLE DEVICES WERE SUCCESSFULLY PRE-PLACED. THE SHEATH WAS UPSIZED TO 24F AND THE MICRA IMPLANT PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURES. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
721172 PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 12773-03 2100741 08717648235184

Patients

Seq Age Sex Outcome Treatment
1 75 YR Male Required Intervention