FDA Adverse Event
Malfunction
Summary report: N
TURBOWHISKER BLADE EP-1; 4.5MM
MDR report key: 1635645
·
Received March 10, 2010
Report
- Report Number
- 1219602-2010-00044
- Event Type
- Malfunction
- Date Received
- March 10, 2010
- Date of Event
- March 17, 2009
- Report Date
- March 19, 2009
- Manufacturer
- MANSFIELD MANUFACTURING SITE
- Product Code
- NBH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A CLEANING SOLUTION HAS BEEN IMPLEMENTED DURING THE METAL FINISHING PREP STEP. THIS HAS ELIMINATED ANY EDM RESIDUE, PRIOR TO ASSEMBLY. (B) (4).
Description of Event or Problem · 1
STARTED CASE AND DOCTOR NOTICED AN OILY SUBSTANCE COMING OUT OF BLADE, OPENED ANOTHER SAME LOT NUMBER AND SAME ISSUE. WENT TO ANOTHER TYPE OF BLADE AND WAS ABLE TO FINISH CASE. FLUSHED OUT OILY SUBSTANCE. METAL SHAVING ALSO AT THE BEGINNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TURBOWHISKER BLADE EP-1; 4.5MM | TURBOWHISKER BLADE EP-1; 4.5MM / NBH | NBH | MANSFIELD MANUFACTURING SITE | 7205316 | 20126463 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |