PREPSTAIN SLIDE PROCESSOR
Report
- Report Number
- 1062336-2010-00003
- Event Type
- Injury
- Date Received
- March 17, 2010
- Date of Event
- February 8, 2010
- Report Date
- March 17, 2010
- Manufacturer
- TRIPATH IMAGING
- Product Code
- KPA
- PMA / PMN Number
- P970018
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ALTHOUGH THEY UTILIZE THE SAME EQUIPMENT, THE NON-GYNECOLOGICAL APPLICATION IS NOT INCLUDED PMA P970018 FOR THE PREPSTAIN SLIDE PROCESSOR. VERIFIED CUSTOMER IS USING SOFTWARE VERSIONS (B) (4) AND UTILITY SOFTWARE (B) (4). THIS SOFTWARE WAS NOT PART OF ANY PREVIOUS FIELD ACTIONS OR RECALLS. OF THE 6 NON-GYNECOLOGICAL SAMPLES PROCESSED THAT DAY, THREE SAMPLES WERE FOUND TO BE MIX-MATCHED; HOWEVER, ONLY ONE RESULTED IN AN ADVERSE PT OUTCOME. TRAY 1 CASE 3 SLIDE- SHOULD HAVE BEEN AN OVARIAN ADENOCARCINOMA IN AN ASCITES FLUID, BUT THE SLIDE CONTAINED BENIGN CELLS ON THE CYTOLOGY. THERE WAS A MALIGNANT CELL BLOCK AND MALIGNANT SURGICAL SPECIMEN FOR OVARIAN CANCER. THE CASE WAS REPORTED AS ADENOCARCINOMA. TRAY 1 CASE 4 SLIDE- SHOULD HAVE BEEN THE BENIGN PANCREATIC CYST FLUID, BUT THE SLIDE CONTAINED ABUNDANT ADENOCARCINOMA FROM A MALIGNANT PANCREATIC CYST FLUID. THERE WAS NOT A CELL BLOCK ON THIS CASE. THE CASE WAS REPORTED AS ADENOCARCINOMA. TRAY 1 CASE 5 SLIDE- SHOULD HAVE BEEN CELLS FROM THE ADENOCARCINOMA OF PANCREAS, BUT CONTAINED ABUNDANT CELLS FROM THE OVARIAN ADENOCARCINOMA. THERE WAS A MALIGNANT CELL BLOCK ON THIS CASE AND THE CASE WAS REPORTED AS ADENOCARCINOMA. THE THREE CASES INVOLVED WERE "WALL TO WALL CANCER" WITH NO EVIDENCE OF INCIDENTAL CONTAMINATION. ON MONDAY, (B) (6), THE LABORATORY RAN A 24-SAMPLE TEST RUN AND VERIFIED THAT TUBE 1 GOES TO SLIDE 1, TUBE 2 TO SLIDE 2, TUBE 3 TO SLIDE 3 THROUGH 24. EVERYTHING WAS DELIVERED CORRECTLY. THEY MONITORED SPECIMEN DELIVERY FOR 2 DAYS AND CONCLUDED THAT THIS IS NOT A CASE OF MACHINE CROSS CONTAMINATION. BOTH THE PREP TECH THAT PROCESSED THE SAMPLES AND THE CYTOLOGY SUPERVISOR THINK THAT THIS WAS AN OPERATOR ERROR.
ON MONDAY, (B) (6) 2010, 6 NON-GYNECOLOGICAL CASES WERE PROCESSED BY A PREPARATION TECHNOLOGIST THAT DID NOT USUALLY PROCESS ON THE PREPSTAIN SLIDE PROCESSOR. THE CYTOLOGY SUPERVISOR REPORTED THAT THE DAY WAS VERY BUSY AND THE PREP TECH DID NOT VERIFY THAT THE SLIDES AND TUBES MATCHED. ON THURSDAY, (B) (6) 2010, A PT WITH A REPORTED PANCREATIC ADENOCARCINOMA PROCESSED ON (B) (6) HAD THE TAIL SECTION OF HER PANCREAS REMOVED AND PART OF HER STOMACH AND SPLEEN REMOVED AS A RESULT OF THIS DIAGNOSIS. WE DO NOT KNOW WHETHER THE SURGERY WAS DONE PRIMARILY BECAUSE OF THE CYTOLOGY RESULT. HOWEVER, THE PT UNDERWENT A SURGICAL PROCEDURE WITH EVERY EXPECTATION OF FINDING CANCER BASED UPON THE ERRONEOUS CYTOLOGY RESULT. AN OPERATION FOR A PANCREATIC CYST WOULD NOT TYPICALLY HAVE BEEN THIS EXTENSIVE. THE EXTENSIVE NATURE OF THE SURGERY IS USUALLY ONLY DONE AS PART OF A CURATIVE CANCER SURGERY. IF THAT IS THE CASE, THE MIX-UP IN THE CYTOLOGY SPECIMENS CLEARLY HAD AN ADVERSE IMPACT ON THE PT. IT DOES NOT APPEAR FROM ANY OF THE AVAILABLE INFO THAT A MALFUNCTION OF THE PREPSTAIN SYSTEM HAD ANYTHING TO DO WITH THE MIX-UP IN SPECIMENS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PREPSTAIN SLIDE PROCESSOR | NONE | KPA | TRIPATH IMAGING | TECAN AG | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other |