FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM

MDR report key: 16351276 · Received February 10, 2023

Report

Report Number
0002023141-2023-00463
Event Type
Injury
Date Received
February 10, 2023
Date of Event
July 22, 2022
Report Date
May 25, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024020085
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ZIMMERBIOMET COMPLAINT NUMBER CMP ¿ (B)(4). DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1252793. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED.

Additional Manufacturer Narrative · 0

ZIMMERBIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. ADDITIONAL PMA/510(K) NUMBER ¿ K011028, K013227. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATION 3 WAS REMOVED DUE TO PERI-IMPLANTITIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
996939 IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 8MM DENTAL IMPLANT DZE ZIMMER DENTAL TSVWB8 1252793 00889024020085

Patients

Seq Age Sex Outcome Treatment
1 48 YR Female Required Intervention