FDA Adverse Event Malfunction Summary report: N

C5 SYSTEM

MDR report key: 16349701 · Received February 10, 2023

Report

Report Number
9611109-2023-00061
Event Type
Malfunction
Date Received
February 10, 2023
Date of Event
January 12, 2023
Report Date
February 9, 2023
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
DTQ
PMA / PMN Number
K093882
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H10: DURING TROUBLESHOOTING AT CUSTOMER'S FACILITY, COMPUTER BOARD (HKR) WAS FOUND TO BE DEFECTIVE. THEREFORE, IT WAS REPLACED IN ORDER TO SOLVE THE ISSUE. SUBSEQUENT FUNCTIONAL VERIFICATION TESTING WAS COMPLETED WITHOUT FURTHER ISSUES AND THE UNIT WAS RETURNED TO SERVICE. BASED ON TECHNICAL SERVICE INTERVENTION DONE, IT CANNOT BE RULED OUT THAT THE REPORTED ISSUE WAS CAUSED BY AN ELECTRONIC FAILURE OF THE PUMP COMPUTER BOARD (HKR).

Additional Manufacturer Narrative · 0

THERE WAS NO PATIENT INVOLVEMENT. LIVANOVA DEUTSCHLAND MANUFACTURES THE C5 SYSTEM. THE INCIDENT OCCURRED IN NITAI NAGAR, INDIA. A LIVANOVA FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO INVESTIGATE THE DEVICE AND COULD CONFIRM THE REPORTED ISSUE. CONNECTIONS OF MOTOR CONTROL AND MOTOR POWER AMPLIFIER BOARDS WERE CHECKED AND RE-FIXED WITHOUT ISSUE RESOLUTION. THERE WAS NO SIGN OF SPILLAGE ON BOARDS. SOFTWARE OF THE DEVICE WAS FLASHED TO A NEW VERSION, BUT THE ISSUE STILL REMAINED. IN ORDER TO FIX THE ISSUE, MOTOR CONTROL AND MOTOR POWER AMPLIFIER BOARDS NEED TO BE REPLACED. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 0

SEE INITIAL REPORT.

Description of Event or Problem · 0

LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A PUMP ON A C5 HEART LUNG MACHINE GAVE AN ERROR MESSAGE RELATED TO A MOTOR CONTROL FAILED. THE ISSUE OCCURRED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47287 C5 SYSTEM CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS DTQ LIVANOVA DEUTSCHLAND 58-00-00

Patients

Seq Age Sex Outcome Treatment
1 Unknown