FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET

MDR report key: 16349116 · Received February 10, 2023

Report

Report Number
1024879-2023-00080
Event Type
Malfunction
Date Received
February 10, 2023
Date of Event
February 3, 2023
Report Date
June 26, 2023
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903673235
PMA / PMN Number
K030573
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVALUATION: YES D9: RETURNED TO MANUFACTURER ON: 2023-02-06 H.6. INVESTIGATION SUMMARY: "MATERIAL #: 367323 LOT/BATCH #: 2040297 BD RECEIVED 1 SAMPLE FOR INVESTIGATION. THE SAMPLE WAS EVALUATED BY VISUAL EXAMINATION AND THE INDICATED FAILURE MODE FOR BROKEN LUER CAP WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, 100 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF BROKEN LUER CAP WAS OBSERVED IN 1 OF THE SAMPLES. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE BROKEN LUER CAP. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE MEDICAL DEVICE TYPES REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH 510(K) NUMBER IS AS FOLLOWS: \COMMON DEVICE NAME: INTRAVASCULAR ADMINISTRATION SET. MEDICAL DEVICE TYPE: FPA. DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET THERE NON PATIENT NEEDLE SEPARATED FROM THE HOLDER LUER/ADAPTOR/HUB. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "THE MALE LUER PART OF THE CAP BROKE OFF AND GOT STUCK IN THE FEMALE LUER OF THE EXTENSION.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET THERE NON PATIENT NEEDLE SEPARATED FROM THE HOLDER LUER/ADAPTOR/HUB. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "THE MALE LUER PART OF THE CAP BROKE OFF AND GOT STUCK IN THE FEMALE LUER OF THE EXTENSION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
43163 BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 367323 2040297 50382903673235

Patients

Seq Age Sex Outcome Treatment
1 Unknown