UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEMS
MDR report key: 1634881
·
Received March 18, 2010
Report
- Report Number
- 2050012-2010-00121
- Event Type
- Malfunction
- Date Received
- March 18, 2010
- Date of Event
- February 9, 2010
- Report Date
- March 18, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JGS
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
BEFORE THE EVENT, THE ISE SYSTEM WAS CALIBRATED AND NA QC RESULTS WERE WITHIN THE LAB'S ESTABLISHED RANGES. PRE-ANALYTICAL SAMPLE HANDLING WAS DISCUSSED WITH CUSTOMER. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED: THE FSE DECONTAMINATED THE ISE SYSTEM AND COMPLETED A PREVENTIVE MAINTENANCE (PM). A CLEAR ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING ERRONEOUSLY LOW SODIUM (NA) RESULTS GENERATED BY THE UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEMS. THE RESULTS WERE REPORTED OUT OF THE LAB. PATIENT TREATMENT WAS NOT AFFECTED.