BD INTIMA-II CLOSED IV CATHETER SYSTEM
Report
- Report Number
- 3014704491-2023-00011
- Event Type
- Malfunction
- Date Received
- February 10, 2023
- Date of Event
- October 10, 2022
- Report Date
- February 15, 2023
- Manufacturer
- BD (SUZHOU)
- Product Code
- FOZ
- UDI-DI
- 00382903830831
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 2011049. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED WHILE USING BD INTIMA-II CLOSED IV CATHETER SYSTEM THE TUBING WAS SPLIT. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: ADVERSE EVENT STATUS: INDWELLING CATHETER SOFT NEEDLE RUPTURE.
IT WAS REPORTED WHILE USING BD INTIMA-II CLOSED IV CATHETER SYSTEM THE TUBING WAS SPLIT. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: ADVERSE EVENT STATUS: INDWELLING CATHETER SOFT NEEDLE RUPTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 48027 | BD INTIMA-II CLOSED IV CATHETER SYSTEM | INTRAVASCULAR CATHETER | FOZ | BD (SUZHOU) | 2011049 | 00382903830831 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |