FDA Adverse Event Malfunction Summary report: N

PKG, 1488 HD 3-CHIP CAMERA HEAD, C-MOUNT

MDR report key: 16342759 · Received February 9, 2023

Report

Report Number
0002936485-2023-00111
Event Type
Malfunction
Date Received
February 9, 2023
Date of Event
January 23, 2023
Report Date
August 24, 2023
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
GCJ
UDI-DI
07613327061055
PMA / PMN Number
K132785
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ALLEGED FAILURE: BLURRY GRAINY IMAGE. CONFIRMED FAILURE: CHD-A-LCJ LOOSE CABLE JACKET PROBABLE ROOT CAUSE: ¿ CABLES ¿ CONNECTORS ¿ DIGITAL BOARD ¿ POWER SUPPLY ¿ FILTER / FUSE ¿ AC INLET BOARD ¿ MAIN BOARD ¿ TRANSITION BOARD ¿ FIBER BOARD ¿ INTERMITTENCE BETWEEN TRANSITION BOARD AND CH CONNECTOR ¿ SOFTWARE ¿SCOPE (SEE RSK10356 OUTPUT ""VISUALIZATION/IMAGE"") ¿LIGHT SOURCE (SEE RSK10975 OUTPUT ""WHITE LIGHT"") ¿ CAMERA HEAD (SENSOR, SENSOR CABLE) ¿ COUPLER (DUST ON OPTICS, SCRATCHED OPTICS, BROKEN / SHIFTED LENSES, FOCUSING MECHANISM, ENDOCLAMP, ZOOM RING) ¿ 1488 & 1588 EXTENSION CABLE ¿ INTERMITTENCE WHILE USING RF DEVICES (EX: VALLEYLAB) ¿ PROBLEM TOGGLING LIGHT SOURCE OR FROM CAMERA HEAD ¿ CONNECTED MONITORS ¿ LEAKING ¿ USE ERROR ¿ POOR GROUNDING (E.G. ""FINGER"" GROUNDING TAB CONNECTING FRONT AND REAR ENCLOSURE. CAMERA HEAD CONNECTION FROM CABLE TO TRANSITION BOARD.) ¿ELECTROMAGNETIC INTERFERENCE (EMI) FROM RF COMMUNICATION, HF SURGICAL INSTRUMENTS, ESD, OR POWER SURGE ¿ MANUFACTURING/ SERVICE NONCONFORMITY" THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED. THE DEVICE MANUFACTURER DATE IS NOT KNOWN AT THIS TIME. HOWEVER, SHOULD IT BECOME AVAILABLE IT WILL BE PROVIDED IN FUTURE REPORTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS BLURRY IMAGE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS BLURRY IMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1067742 PKG, 1488 HD 3-CHIP CAMERA HEAD, C-MOUNT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ STRYKER ENDOSCOPY-SAN JOSE 1488210105 07613327061055

Patients

Seq Age Sex Outcome Treatment
1 Unknown