Narratives
THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE DETERMINED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION. IF THE DEVICE BECOMES AVAILABLE FOR EVALUATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.
BASED ON THE INFORMATION PROVIDED WHICH MAY INCLUDE THE DEVICE (IF AVAILABLE AND RETURNED), PICTURES, VIDEOS, EVENT DESCRIPTION, AND ANY ADDITIONAL INFORMATION FROM THE FIELD, ARTHREX WAS ABLE TO CONCLUDE A MOST LIKELY CAUSE. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO USER ERROR OF THE DEVICE DUE TO MISALIGNED INSERTION AND/OR PRYING/LEVERAGING DURING USE.
ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT WHILE USING AN AR-9301-03 ARTHREX ECLIPSE CAGE SCREW LARGE, AN AR-9400-SBK ECLIPS SPEEDSCAP IMPLANT SYSTEM AND AN AR-9301-51CPC ECLIPSE TRUNION, DURING THE FINAL STEPS, THE SURGEON DRILLED THE PILOT HOLE INTO THE SCREW AND FIXATION WAS COMPROMISED. SURGEON MOVED TO AN APEX SYSTEM INSTEAD. WHILE USING THE APEX, FIXATION WAS ALSO COMPROMISED WHEN TRAILING THE HEAD, SO SURGEON USED A LARGER SIZE INSTEAD. ADDITIONAL INFORMATION REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1607217 | ECLIPSE TRUNION 51MM, SLOTTED, TPS CAP | TOTAL SHOULDER ARTHROPLASTY SYSTEM | QHQ | ARTHREX, INC. | ECLIPSE TRUNION 51MM, SLOTTED, TPS CAP | 19.04146 | 00888867276185 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |