FDA Adverse Event Injury Summary report: N

PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM

MDR report key: 16336163 · Received February 9, 2023

Report

Report Number
2024168-2023-01337
Event Type
Injury
Date Received
February 9, 2023
Date of Event
January 17, 2023
Report Date
February 24, 2023
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
UDI-DI
08717648235184
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE ADDITIONAL PROSTYLE DEVICES WITH REPORTED ISSUES REFERENCED ARE FILED UNDER SEPARATE MEDWATCH REPORT NUMBERS.

Additional Manufacturer Narrative · 0

ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D4: LOT NUMBER CHANGED FROM LOT #2102042 TO LOT #2102441.

Description of Event or Problem · 0

IT WAS REPORTED THIS WAS AN ARTERIOTOMY CLOSURE OF THE LEFT COMMON FEMORAL ARTERY USING THE PRE-CLOSE TECHNIQUE VIA A 6F SHEATH HOLE PRIOR TO A TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) INTERVENTIONAL PROCEDURE. REPORTEDLY, A CUFF MISS [SUTURE RETRIEVAL ISSUE] WAS NOTICED WITH FOUR PROSTYLE DEVICES. THE SUTURES OF TWO NEW PROSTYLE DEVICES WERE SUCCESSFULLY PRE-PLACED. THE SHEATH WAS UPSIZED TO A 16F SHEATH AND THE TAVR PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURES. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1491345 PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 12773-03 2102441 08717648235184

Patients

Seq Age Sex Outcome Treatment
1 78 YR Male Required Intervention