PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM
Report
- Report Number
- 2024168-2023-01337
- Event Type
- Injury
- Date Received
- February 9, 2023
- Date of Event
- January 17, 2023
- Report Date
- February 24, 2023
- Manufacturer
- ABBOTT VASCULAR
- Product Code
- MGB
- UDI-DI
- 08717648235184
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE ADDITIONAL PROSTYLE DEVICES WITH REPORTED ISSUES REFERENCED ARE FILED UNDER SEPARATE MEDWATCH REPORT NUMBERS.
ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D4: LOT NUMBER CHANGED FROM LOT #2102042 TO LOT #2102441.
IT WAS REPORTED THIS WAS AN ARTERIOTOMY CLOSURE OF THE LEFT COMMON FEMORAL ARTERY USING THE PRE-CLOSE TECHNIQUE VIA A 6F SHEATH HOLE PRIOR TO A TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) INTERVENTIONAL PROCEDURE. REPORTEDLY, A CUFF MISS [SUTURE RETRIEVAL ISSUE] WAS NOTICED WITH FOUR PROSTYLE DEVICES. THE SUTURES OF TWO NEW PROSTYLE DEVICES WERE SUCCESSFULLY PRE-PLACED. THE SHEATH WAS UPSIZED TO A 16F SHEATH AND THE TAVR PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURES. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1491345 | PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM | SUTURE MEDIATED CLOSURE | MGB | ABBOTT VASCULAR | 12773-03 | 2102441 | 08717648235184 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Male | Required Intervention |