FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1,13,MTX,MG

MDR report key: 16334070 · Received February 9, 2023

Report

Report Number
0002023141-2023-00479
Event Type
Injury
Date Received
February 9, 2023
Report Date
March 31, 2023
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019843
PMA / PMN Number
K101977
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: H6: TYPE OF INVESTIGATION CODES WERE UPDATED: 3331, 4109, 4110, 4111. H6: INVESTIGATION FINDINGS CODE WAS UPDATED: 213. H6: INVESTIGATION CONCLUSION CODE WAS UPDATED: 4310. ONE (1) IMP,TSV,4.1,13,MTX,MG (TSVT4B13) WAS RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED IMPLANT IDENTIFIED SIGNS OF USE; BONE DEBRIS PRESENT ON EXTERNAL THREADS. THERE WAS NO DAMAGE, WEAR, OR SIGNS OF MALFUNCTION THAT WOULD CONTRIBUTE TO THE EVENT. THE IMPLANT MATCHED PRINT SPECIFICATIONS. BASED ON THE EVALUATION, DEVICE MALFUNCTION HAS NOT OCCURRED. ADDITIONALLY, THERE IS NO EXISTING NONCONFORMANCE / CAPA / HHE/D / IE / PRODUCT HOLD AGAINST THE REPORTED DEVICE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. MONTHLY POST MARKET TRENDING REVIEW IDENTIFIED NO ACTIONABLE TRENDS OR CORRECTIVE ACTIONS FOR THE REPORTED EVENT OR DEVICE. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. THEREFORE, BASED ON THE AVAILABLE INFORMATION, THE PRODUCT WAS WITHIN SPECIFICATIONS AND CONFORMING WHEN IT LEFT ZIMVIE. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1254007. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1254007 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. FUNCTIONAL TESTING COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICES AND EVENT. THEREFORE, THE REPORTED EVENT COULD NOT BE RECREATED. THE COMPLAINT IS NOT RELATED TO THE FUNCTIONAL PERFORMANCE OF THE DEVICE. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, BASED ON THE INVESTIGATION AND RISK FILE REVIEW, THE MOST LIKELY CAUSE DETERMINED FROM THE INVESTIGATION IS PATIENT SPECIFIC ISSUE, IMPLANT NOT PLACED PROPERLY, CLINICIAN ERROR AND INADEQUATE TREATMENT PLANNING. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER: (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. DATE OF EVENT UNKNOWN / NOT PROVIDED. PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION WAS RECEIVED AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT ON TOOTH SITE #31 WAS REMOVED DUE TO NUMBNESS AND PAIN AFTER THE DAY OF SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1817194 IMP,TSV,4.1,13,MTX,MG DENTAL IMPLANT DZE ZIMMER DENTAL TSVT4B13 1254007 00889024019843

Patients

Seq Age Sex Outcome Treatment
1 63 YR Female Required Intervention