FDA Adverse Event Injury Summary report: N

MAQUET

MDR report key: 1633001 · Received March 12, 2010

Report

Report Number
9710055-2010-00006
Event Type
Injury
Date Received
March 12, 2010
Date of Event
February 11, 2010
Report Date
February 11, 2010
Manufacturer
MAQUET S.A.
Product Code
FSY
PMA / PMN Number
K070442
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE LED LIGHT HEAD IS SECURED TO THE MAIN ARM VIA A RETAINING CLIP. THIS CLIP IS KEPT IN PLACE BY A PROTECTIVE COVER/SLEEVE WHICH IN TURN IS SECURED TO THE LIGHTING UNIT BY A SCREW TO PREVENT ITS MOVEMENT. MANUFACTURER ANALYSIS FOUND THAT THE SAFETY COVER WAS NOT INSTALLED PROPERLY ON THIS LIGHTING UNIT. THE COVER WAS MISSING ITS FASTENING SCREW. BECAUSE THIS SCREW WAS MISSING, THE COVER WAS ABLE TO SLIDE OUT OF POSITION AND THE RETAINING CLIP WORKED ITS WAY LOOSE ALLOWING THE CUPOLA ASSEMBLY TO FALL. TO PREVENT THIS TYPE OF OCCURRENCE, MAQUET PLACES YELLOW LABELS DIRECTLY ON THE SPRING ARMS STATING THAT "THE SLEEVE MUST BE SECURED BY SCREW". ADDITIONAL INSTRUCTIONS AND WARNINGS ARE ALSO INCLUDED IN THE POWERLED INSTALLATION AND SERVICE MANUALS. MAQUET (B) (4) HAS FACILITATED A REPAIR OF THIS LIGHT ENSURING THAT THE LOCKING PIN IS IN PLACE AND THAT THE RETAINING SLEEVE IS SECURED PROPERLY IN CORRECT POSITION. MAQUET INC. SUBMITS THIS REPORT ON BEHALF OF THE DEVICE MANUFACTURING FACILITY. MAQUET S.A. PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT.

Description of Event or Problem · 1

CLEANING STAFF MEMBER WAS CLEANING THE OPERATING ROOM LIGHTS. WHEN SHE PULLED THE CUPOLA HANDLE TOWARDS HER TO CLEAN THE CUPOLA, THE CUPOLA DISENGAGED FROM THE SPRING ARM, AND STRUCK THE CLEANING STAFF MEMBER MID-FACE. THE CUPOLA THEN LANDED ON THE FLOOR. THE CUPOLA FRACTURED THE STAFF MEMBER'S NOSE, AND BROKE HER EYEGLASSES. (B) (4)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAQUET FSY MAQUET S.A. POWERLED

Patients

Seq Age Sex Outcome Treatment
1 Other