FDA Adverse Event Injury Summary report: N

LINER: MPACT FLAT PE HC LINER Ø36/F

MDR report key: 16322503 · Received February 8, 2023

Report

Report Number
3005180920-2023-00042
Event Type
Injury
Date Received
February 8, 2023
Date of Event
January 13, 2023
Report Date
February 8, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LPH
UDI-DI
07630030812132
PMA / PMN Number
K103721
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 18 JANUARY 2023: LOT 186971: 108 ITEMS MANUFACTURED AND RELEASED ON 19-OCT-2018. EXPIRATION DATE: 2023-10-04. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 103 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ANOTHER SIMILAR REPORTED EVENT IN THE PERIOD OF REVIEW. ADDITIONAL INVOLVED IMPLANT: BATCH REVIEW PERFORMED ON 18 JANUARY 2023 ON BALL HEADS: MECTACER 01.29.208 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE S - 4 (K112115) LOT. 2211903 LOT 2211903: 200 ITEMS MANUFACTURED AND RELEASED ON 25-AUG-2022. EXPIRATION DATE: 2027-07-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 160 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT IN THE PERIOD OF REVIEW.

Description of Event or Problem · 0

THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. AT ABOUT 1 MONTH FROM PRIMARY THE SURGEON PERFORMED A WASHOUT AND REVISED THE POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2314002 LINER: MPACT FLAT PE HC LINER Ø36/F HIP HIGH CROSS POLYETHYLENE LINER LPH MEDACTA INTERNATIONAL SA 01.32.3648HCT 186971 07630030812132

Patients

Seq Age Sex Outcome Treatment
1 Male Required Intervention