FDA Adverse Event
Injury
Summary report: N
AMSCO
MDR report key: 163215
·
Received April 16, 1998
Report
- Report Number
- 163215
- Event Type
- Injury
- Date Received
- April 16, 1998
- Date of Event
- December 24, 1997
- Report Date
- March 6, 1998
- Manufacturer
- AMSCO CORP QA
- Product Code
- LGZ
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
A 5 LITER PLASTIC BAG OF SALINE IRRIGANT WARMED IN WARMING CABINET USED AS POSITIONING DEVICE DURING SURGERY. PT REC'D 3RD DEGREE BURN REQUIRING SKIN GRAFTING. TEMP MEASUREMENTS INSIDE THE CABINETS RANGED FROM 38.1 C TO 83.1 C WITH AVG READINGS RANGING FROM 39.8 TO 63.8 C. WARMING CABINETS WERE SET AT 170 F.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AMSCO | WARMING CABINETS | LGZ | AMSCO CORP QA | M70 WC-EL | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Hospitalization| R |