VITROS CHEMISTRY PRODUCTS BENZ REAGENT
Report
- Report Number
- 1319808-2010-00002
- Event Type
- Malfunction
- Date Received
- February 5, 2010
- Date of Event
- December 24, 2009
- Report Date
- January 8, 2010
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- JXM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION INTO THIS EVENT CONCLUDED THAT A FALSE NEGATIVE BENZ RESULT OCCURRED. THE MOST LIKELY CAUSE IF A COMBINATION OF A KNOWN LIMITATION OF THE VITROS BENZ REAGENT RELATED TO LOW CROSS REACTIVITY WITH LORAZEPAM GLUCURONIDE AND USE ERROR IN THE INTERPRETATION OF A VITROS BENZ RESULT. CONFIRMATORY TESTING CONCLUDES THAT THE VITROS BENZ METHOD IDENTIFIED A BENZODIAZEPINE COMPOUND AT A MUCH LOWER CONCENTRATION THAN GC/MS. THE INTENDED USE SECTION OF THE VITROS BENZ IFU STATES THAT THE VITROS CHEMISTRY PRODUCT BENZ ASSAY IS INTENDED FOR USE BY PROFESSIONAL LABORATORY PERSONNEL. IT PROVIDED ONLY A PRELIMINARY TEST RESULT. A MORE SPECIFIC ALTERNATIVE CHEMICAL METHOD MUST BE USED TO CONFIRM A RESULT OBTAINED WITH VITROS BENZ ASSAY. GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) IS THE PREFERRED CONFIRMATORY METHOD. CLINICAL CONSIDERATION AND PROFESSIONAL JUDGEMENT SHOULD BE APPLIED TO ANY DRUG-OF-ABUSE TEST RESULT. NO MALFUNCTION OF THE VITROS BENZ REAGENT OCCURRED. THERE WAS NO ALLEGATION OF PT HARM AS A RESULT OF THIS EVENT.
A CUSTOMER OBSERVED FALSE NEGATIVE VITROS BENZ RESULT ON ONE PT SAMPLE WHILE USING THE VITROS 5, 1 FS CHEMISTRY SYSTEM. THE MAGNITUDE AND DIRECTION OF THE FALSE NEGATIVE VITROS BENZ RESULT OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION. THE ORIGINAL AFFECTED RESULT WAS REPORTED TO THE CLINICIAN, AND A CORRECTED REPORT WAS LATER ISSUED. THERE WAS NO REPORT OF HARM TO THE PT AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITROS CHEMISTRY PRODUCTS BENZ REAGENT | IN-VITRO DIAGNOSTIC | JXM | ORTHO-CLINICAL DIAGNOSTICS | NA | 1523-07-9075 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |