FDA Adverse Event Malfunction Summary report: N

VITROS CHEMISTRY PRODUCTS BENZ REAGENT

MDR report key: 1631970 · Received February 5, 2010

Report

Report Number
1319808-2010-00002
Event Type
Malfunction
Date Received
February 5, 2010
Date of Event
December 24, 2009
Report Date
January 8, 2010
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
JXM
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION INTO THIS EVENT CONCLUDED THAT A FALSE NEGATIVE BENZ RESULT OCCURRED. THE MOST LIKELY CAUSE IF A COMBINATION OF A KNOWN LIMITATION OF THE VITROS BENZ REAGENT RELATED TO LOW CROSS REACTIVITY WITH LORAZEPAM GLUCURONIDE AND USE ERROR IN THE INTERPRETATION OF A VITROS BENZ RESULT. CONFIRMATORY TESTING CONCLUDES THAT THE VITROS BENZ METHOD IDENTIFIED A BENZODIAZEPINE COMPOUND AT A MUCH LOWER CONCENTRATION THAN GC/MS. THE INTENDED USE SECTION OF THE VITROS BENZ IFU STATES THAT THE VITROS CHEMISTRY PRODUCT BENZ ASSAY IS INTENDED FOR USE BY PROFESSIONAL LABORATORY PERSONNEL. IT PROVIDED ONLY A PRELIMINARY TEST RESULT. A MORE SPECIFIC ALTERNATIVE CHEMICAL METHOD MUST BE USED TO CONFIRM A RESULT OBTAINED WITH VITROS BENZ ASSAY. GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) IS THE PREFERRED CONFIRMATORY METHOD. CLINICAL CONSIDERATION AND PROFESSIONAL JUDGEMENT SHOULD BE APPLIED TO ANY DRUG-OF-ABUSE TEST RESULT. NO MALFUNCTION OF THE VITROS BENZ REAGENT OCCURRED. THERE WAS NO ALLEGATION OF PT HARM AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

A CUSTOMER OBSERVED FALSE NEGATIVE VITROS BENZ RESULT ON ONE PT SAMPLE WHILE USING THE VITROS 5, 1 FS CHEMISTRY SYSTEM. THE MAGNITUDE AND DIRECTION OF THE FALSE NEGATIVE VITROS BENZ RESULT OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION. THE ORIGINAL AFFECTED RESULT WAS REPORTED TO THE CLINICIAN, AND A CORRECTED REPORT WAS LATER ISSUED. THERE WAS NO REPORT OF HARM TO THE PT AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITROS CHEMISTRY PRODUCTS BENZ REAGENT IN-VITRO DIAGNOSTIC JXM ORTHO-CLINICAL DIAGNOSTICS NA 1523-07-9075

Patients

Seq Age Sex Outcome Treatment
1