FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ SYRINGE 20 ML BULK STERILE PHARMACY CONVENIENCE TRAY

MDR report key: 16317090 · Received February 7, 2023

Report

Report Number
9610847-2023-00016
Event Type
Malfunction
Date Received
February 7, 2023
Date of Event
January 16, 2023
Report Date
March 17, 2023
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMF
UDI-DI
00382903056170
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES, D10: RETURNED TO MANUFACTURER ON: 24-JAN-2023. H6: INVESTIGATION SUMMARY: OUR QUALITY ENGINEER INSPECTED THE 1 PHOTO AND 1 SAMPLE SUBMITTED FOR EVALUATION. THE REPORTED ISSUE OF FOREIGN MATTER WAS CONFIRMED UPON INSPECTION OF THE SAMPLE. ANALYSIS OF THE PHYSICAL SAMPLE SHOWED THAT THERE WAS A FOREIGN MATTER PARTICULATE ON THE RETURNED SAMPLE. FTIR TESTING WAS PERFORMED IN AN ATTEMPT TO DETERMINE THE COMPOSITION OF THE MATERIAL, BUT THE RESULT OF THE TEST WAS INCONCLUSIVE. SINCE THE FTIR RESULTS WERE INCONCLUSIVE, AND AN EXACT MATERIAL COMPOSITION COULD NOT BE DETERMINED, AN EXACT MANUFACTURING RELATED ROOT CAUSE COULD NOT BE DETERMINED. PRODUCTION RECORDS WERE REVIEWED, AND THIS BATCH MET OUR MANUFACTURING PRODUCT SPECIFICATION REQUIREMENTS.

Additional Manufacturer Narrative · 0

DATE OF EVENT IS UNKNOWN; AWARENESS DATE HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD LUER-LOK¿ SYRINGE 20 ML BULK STERILE PHARMACY CONVENIENCE TRAY FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONTAMINANTS AS REPORTED BY OUR CUSTOMER. 20 ML, LOT # - 2123361.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD LUER-LOK¿ SYRINGE 20 ML BULK STERILE PHARMACY CONVENIENCE TRAY FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONTAMINANTS AS REPORTED BY OUR CUSTOMER. 20 ML, LOT # - 2123361.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2229409 BD LUER-LOK¿ SYRINGE 20 ML BULK STERILE PHARMACY CONVENIENCE TRAY PISTON SYRINGE FMF BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 2123361 00382903056170

Patients

Seq Age Sex Outcome Treatment
1 Unknown