ON-BOARD IMAGER (OBI)
Report
- Report Number
- 2916710-2009-00083
- Event Type
- Injury
- Date Received
- March 11, 2010
- Date of Event
- November 9, 2009
- Report Date
- November 11, 2009
- Manufacturer
- VARIAN MEDICAL SYSTEMS
- Product Code
- IYE
- PMA / PMN Number
- K042720
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
BASED ON THE INITIAL REPORT, VARIAN'S MEDICAL ADVISOR WAS NOT ABLE TO RENDER A MEDICAL EVALUATION ON WHETHER THIS EVENT MAY HAVE CAUSED OR CONTRIBUTED TO A SERIOUS INJURY. VARIAN FIELD STAFF WORKED WITH THE USER TO REPRODUCE THE ISSUE AND THAT INVESTIGATION HAS CONCLUDED THAT THE INCIDENT WAS CAUSED BY USER ERROR - THE NEW DRR WAS FOR A PLAN AT A DIFFERENT ISOCENTER POSITION, AND ATTACHING THIS DRR TO THE PLAN CAUSED THE ISOCENTER SHIFT ERROR. VARIAN WILL FURTHER INVESTIGATE THE ISSUE TO DETERMINE IF THE CAUSE COULD BE MITIGATED BY A CHANGE IN PRODUCT DESIGN. NO ADDITIONAL FOLLOW-UP TO THIS MDR IS EXPECTED.
THE CUSTOMER REPORTED THAT THE STAFF IDENTIFIED AN INCORRECT ISOCENTER SHIFT AFTER PROSTATE PT WAS TREATED FOR 13 OF 39 TREATMENT FRACTIONS. WHEN RE-VISITING THE CONE BEAM CT SCAN (CBCT), IT WAS DISCOVERED THAT A NEW RT LAT ISO-CHECK "DRR" (DIGITALLY RECONSTRUCTED RADIOGRAPH) HAD BEEN CREATED BETWEEN FRACTION 1 AND 2 TO SHOW THE PROSTHETIC HIP. THIS DRR WAS FOR A FIELD AT A DIFFERENT ISOCENTER, BUT WAS APPLIED TO THE ORIGINAL PLAN. THE RESULT WAS AN INCORRECT SHIFT OF THE PLAN 1 CM INFERIOR AND 1 CM ANTERIOR TO THE INTENDED LOCATION. THE TREATING ONCOLOGIST HAS INDICATED THAT DUE TO THIS ISSUE, THERE IS A POTENTIAL OF TREATMENT UNDER DOSE. HOWEVER, IT IS IMPOSSIBLE TO ASCERTAIN THE IMPACT OF POTENTIAL UNDER DOSE. THERE MAY BE A REDUCED PROBABILITY OF TUMOR CONTROL. NO ADDITIONAL PT DATA WAS PROVIDED IN THE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ON-BOARD IMAGER (OBI) | MEDICAL PARTICAL CHARGED RAD. THERAPY | IYE | VARIAN MEDICAL SYSTEMS | H08 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |