FDA Adverse Event Malfunction Summary report: N

HAMILTON MEDICAL AG

MDR report key: 16310365 · Received February 7, 2023

Report

Report Number
3001421318-2023-00086
Event Type
Malfunction
Date Received
February 7, 2023
Date of Event
January 16, 2023
Report Date
May 24, 2023
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
UDI-DI
07630002808590
PMA / PMN Number
K201658
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
505

Narratives

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG COMES TO THE CONCLUSION: INSPIRATORY PORT IS TOO LOOSEN: NOT ALL INFORMATION IS AVAILABLE AT THE TIME OF THIS EVALUATION. REPORTEDLY, INTERVENTION TOOK PLACE, NO PATIENT WAS HARMED. INVESTIGATION WAS INITIATED. ROOT CAUSE: INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG COMES TO THE CONCLUSION: THE CASE CAME TO LIGHT DURING THE ANNUAL MAINTENANCE. THE AFFECTED FOAM PARTS WERE REPLACED AS SPARE PARTS, WHICH SOLVED THE PROBLEM.

Description of Event or Problem · 0

FOLGENDES WURDE GEMELDET AN HAMILTON MEDICAL AG: BEI DIESEM UND MINDESTENS EINEM ZWEITEN GERÄT WIRD GEMELDET, DASS DER INSPIRATIONSANSCHLUSSSTUTZEN LOCKER IST. BEIM AUFSTECKEN UND ABSTECKEN DES INSPIRATIONSSCHLAUCHS WIRD OFFENSICHTLICH DIE BODENFORM AN: 160671 DERART VERFORMT, DASS DIE KUNSTSTOFFPLATTE DES INSPIRATIONSANSCHLUSSES NICHT MEHR AUSREICHEND HALT FINDET. DIES KANN AUCH ZUM ABLÖSEN DES SCHLAUCHES ZWISCHEN DEM BLOWER UND DEM SCHLAUCHANSCHLUSSADAPTER INSPIRATIONSSEITIG FÜHREN. THE FOLLOWING HAS BEEN REPORTED TO HAMILTON MEDICAL AG: ON THIS AND AT LEAST ONE SECOND DEVICE, IT IS REPORTED THAT THE INSPIRATION CONNECTION PIECE IS LOOSE. WHEN ATTACHING AND DETACHING THE INSPIRATION TUBE, THE BOTTOM MOULD AN: 160671 IS APPARENTLY DEFORMED IN SUCH A WAY THAT THE PLASTIC PLATE OF THE INSPIRATION CONNECTOR NO LONGER FINDS SUFFICIENT SUPPORT. THIS CAN ALSO LEAD TO DETACHMENT OF THE HOSE BETWEEN THE BLOWER AND THE HOSE CONNECTION ADAPTER ON THE INSPIRATION SIDE. TRANSLATED WITH WWW.DEEPL.COM/TRANSLATOR (FREE VERSION).

Description of Event or Problem · 0

FOLGENDES WURDE GEMELDET AN HAMILTON MEDICAL AG: BEI DIESEM UND MINDESTENS EINEM ZWEITEN GERÄT WIRD GEMELDET, DASS DER INSPIRATIONSANSCHLUSSSTUTZEN LOCKER IST. BEIM AUFSTECKEN UND ABSTECKEN DES INSPIRATIONSSCHLAUCHS WIRD OFFENSICHTLICH DIE BODENFORM AN: 160671 DERART VERFORMT, DASS DIE KUNSTSTOFFPLATTE DES INSPIRATIONSANSCHLUSSES NICHT MEHR AUSREICHEND HALT FINDET. DIES KANN AUCH ZUM ABLÖSEN DES SCHLAUCHES ZWISCHEN DEM BLOWER UND DEM SCHLAUCHANSCHLUSSADAPTER INSPIRATIONSSEITIG FÜHREN. THE FOLLOWING HAS BEEN REPORTED TO HAMILTON MEDICAL AG: ON THIS AND AT LEAST ONE SECOND DEVICE, IT IS REPORTED THAT THE INSPIRATION CONNECTION PIECE IS LOOSE. WHEN ATTACHING AND DETACHING THE INSPIRATION TUBE, THE BOTTOM MOULD AN: 160671 IS APPARENTLY DEFORMED IN SUCH A WAY THAT THE PLASTIC PLATE OF THE INSPIRATION CONNECTOR NO LONGER FINDS SUFFICIENT SUPPORT. THIS CAN ALSO LEAD TO DETACHMENT OF THE HOSE BETWEEN THE BLOWER AND THE HOSE CONNECTION ADAPTER ON THE INSPIRATION SIDE. TRANSLATED WITH WWW.DEEPL.COM/TRANSLATOR (FREE VERSION)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
107187 HAMILTON MEDICAL AG HAMILTON-C6 VENTILATOR CBK HAMILTON MEDICAL AG 160001 07630002808590

Patients

Seq Age Sex Outcome Treatment
1 Unknown