FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 16302120 · Received February 5, 2023

Report

Report Number
9610877-2023-51125
Event Type
Malfunction
Date Received
February 5, 2023
Date of Event
January 18, 2023
Report Date
February 6, 2023
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDF
UDI-DI
04961333234066
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EC34-I10TL-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K180285. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE WAS BROKEN. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE LG PRONG TOP LOOSENED; HOWEVER, IT IS NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
107263 PENTAX VIDEO COLONOSCOPE FDF HOYA CORPORATION PENTAX TOKYO OFFICE EC34-I10TL 04961333234066

Patients

Seq Age Sex Outcome Treatment
1 Unknown