BD SYRINGE 5 ML S/T BNS
Report
- Report Number
- 1213809-2023-00059
- Event Type
- Malfunction
- Date Received
- February 4, 2023
- Date of Event
- January 10, 2023
- Report Date
- March 2, 2023
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- PMA / PMN Number
- K980987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INITIAL REPORTER PHONE #: (B)(6). A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL, YES. D10: RETURNED TO MANUFACTURER ON: 14-FEB-2023. TEN SAMPLES AND ONE PHOTO WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. A VISUAL INSPECTION WAS PERFORMED, AND TEN LOOSE SAMPLES HAVE LIGHT AND MISSING PRINT OF THE SCALE MARKINGS. THE IMAGE SHOWS SIX LOOSE SYRINGES SHOWING THE NUMBERS AND THE ML AREA OF THE BARREL WITH THE "M" BLURRY ON ALL THE SYRINGES AND THE "L" MISSING ON FOUR SYRINGES. POTENTIAL ROOT CAUSE FOR THE MISSING AND SMEARED PRINT DEFECTS IS ASSOCIATED WITH THE MARKING PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2160605. A REVIEW SHOWED ONE NOTIFICATION DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECTS.
IT WAS REPORTED THAT 327 BD SYRINGE 5 ML S/T BNS EXPERIENCED SCALE MARKING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM SPANISH TO ENGLISH: MANUFACTURING REPORTED TO FIND 327 PARTS OF MATERIAL K-05600-025A BATCH 13P22K0646 WITH A SMEARED PRINTING.
IT WAS REPORTED THAT 327 BD SYRINGE 5 ML S/T BNS EXPERIENCED SCALE MARKING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM SPANISH TO ENGLISH: MANUFACTURING REPORTED TO FIND 327 PARTS OF MATERIAL K-05600-025A BATCH 13P22K0646 WITH A SMEARED PRINTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 81277 | BD SYRINGE 5 ML S/T BNS | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 2160605 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |