FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 1630087 · Received March 12, 2010

Report

Report Number
2954323-2010-00303
Event Type
Malfunction
Date Received
March 12, 2010
Date of Event
February 18, 2010
Report Date
March 12, 2010
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS IS A FINAL REPORT. THE DEVICE ((B) (4)) AND TEST STRIPS (LOT NO: 0934428) HAVE BEEN RETURNED AND AN INVESTIGATION UTILIZING THE RETURNED METER, RETURNED TEST STRIPS AND RETAINED CONTROL SOLUTIONS (LOTS: 9F3P11 AND 9F1P11) HAS DETERMINED THE COMPLAINT IS NOT CONFIRMED. ALL RESULTS WERE WITHIN RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE RESULTS OF "LO" AND 113 MG/DL WERE FOUND IN THE METER'S INTERNAL MEMORY LOG.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON HIS FREESTYLE LITE BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF "LO" (LESS THAN 20 MG/DL) AND 113 MG/DL WITHIN 10 MINUTES. ALL TESTS WERE PERFORMED ON THE FINGER. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 0934428