FREESTYLE LITE
Report
- Report Number
- 2954323-2010-00303
- Event Type
- Malfunction
- Date Received
- March 12, 2010
- Date of Event
- February 18, 2010
- Report Date
- March 12, 2010
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THIS IS A FINAL REPORT. THE DEVICE ((B) (4)) AND TEST STRIPS (LOT NO: 0934428) HAVE BEEN RETURNED AND AN INVESTIGATION UTILIZING THE RETURNED METER, RETURNED TEST STRIPS AND RETAINED CONTROL SOLUTIONS (LOTS: 9F3P11 AND 9F1P11) HAS DETERMINED THE COMPLAINT IS NOT CONFIRMED. ALL RESULTS WERE WITHIN RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE RESULTS OF "LO" AND 113 MG/DL WERE FOUND IN THE METER'S INTERNAL MEMORY LOG.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON HIS FREESTYLE LITE BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF "LO" (LESS THAN 20 MG/DL) AND 113 MG/DL WITHIN 10 MINUTES. ALL TESTS WERE PERFORMED ON THE FINGER. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 0934428 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |