RS GLENOID PLATE EXT CAG +10MM CAGE PEG
Report
- Report Number
- 1038671-2023-00095
- Event Type
- Injury
- Date Received
- February 3, 2023
- Date of Event
- January 1, 2023
- Report Date
- May 14, 2026
- Manufacturer
- EXACTECH, INC.
- Product Code
- KWT
- UDI-DI
- 10885862186706
- PMA / PMN Number
- UNK
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(H3) THE GLENOID BASEPLATE LOOSENING REPORTED WAS LIKELY THE RESULT OF AN INSUFFICIENT BOND BETWEEN THE GLENOID BASEPLATE AND THE BONE AND/OR PATIENT-RELATED CONDITIONS. HOWEVER, THIS CANNOT BE CONFIRMED AS THE DEVICES AND RADIOGRAPHS WERE NOT AVAILABLE FOR EVALUATION. THE FOLLOWING SECTIONS HAVE CORRECTED INFORMATION: (B5) DESCRIBE EVENT OR PROBLEM: IT WAS REPORTED BY THE EQUINOXE SHOULDER STUDY, THAT THIS 55 Y/O MALE PATIENT EXPERIENCED ASEPTIC GLENOID LOOSENING. AT 2-YEAR FOLLOW-UP VISIT X-RAYS REVEALED GLENOID LOOSENING; HOWEVER, PATIENT REPORTS NO PAIN IN RIGHT SHOULDER. THE CASE REPORT FORM INDICATES THIS EVENT IS POSSIBLY RELATED TO DEVICES AND/OR PROCEDURE. THIS EVENT REPORT WAS RECEIVED THROUGH CLINICAL DATA COLLECTION ACTIVITIES. NO ACTION TAKEN, OUTCOME IS CONTINUING. (D1) BRAND NAME: RS GLENOID PLATE EXT CAG +10MM CAGE PEG, PRODUCT CODE: KWT, COMMON DEVICE NAME: PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED (D4) CATALOG NUMBER: 320-15-06, UNIQUE IDENTIFIER (UDI) #: (B)(4). (D10) CONCOMITANT DEVICE(S): 300-01-14 - EQUINOXE HUMERAL STEM PRIMARY PRESS FIT 14MM. 320-38-00 - EQUINOXE REVERSE 38MM HUMERAL LINER +0. 320-10-00 - EQUINOXE REVERSE TRAY ADAPTER PLATE TRAY +0. 320-06-38 - GLENOSPHERE 38MM.
PENDING EVALUATION : CONCOMITANT MEDICAL PRODUCTS : 300-01-14 - EQUINOXE HUMERAL STEM PRIMARY PRESS FIT 14MM; 320-38-00 - EQUINOXE REVERSE 38MM HUMERAL LINER +0; 320-10-00 - EQUINOXE REVERSE TRAY ADAPTER PLATE TRAY +0; 320-15-06 - RS GLENOID PLATE EXT CAG +10MM CAGE PEG.
IT WAS REPORTED BY THE EQUINOXE SHOULDER STUDY, THAT THIS 55 Y/O MALE PATIENT EXPERIENCED ASEPTIC GLENOID LOOSENING. AT 2-YEAR FOLLOW-UP VISIT X-RAYS REVEALED GLENOID LOOSENING; HOWEVER, PATIENT REPORTS NO PAIN IN RIGHT SHOULDER. THE CASE REPORT FORM INDICATES THIS EVENT IS POSSIBLY RELATED TO DEVICES AND/OR PROCEDURE. THIS EVENT REPORT WAS RECEIVED THROUGH CLINICAL DATA COLLECTION ACTIVITIES. NO ACTION TAKEN, OUTCOME IS CONTINUING.
IT WAS REPORTED BY THE EQUINOXE SHOULDER STUDY, THAT THIS 55 Y/O MALE PATIENT EXPERIENCED ASEPTIC GLENOID LOOSENING. AT 2-YEAR FOLLOW-UP VISIT X-RAYS REVEALED GLENOID LOOSENING; HOWEVER, PATIENT REPORTS NO PAIN IN RIGHT SHOULDER. THE CASE REPORT FORM INDICATES THIS EVENT IS POSSIBLY RELATED TO DEVICES AND/OR PROCEDURE. THIS EVENT REPORT WAS RECEIVED THROUGH CLINICAL DATA COLLECTION ACTIVITIES. NO ACTION TAKEN, OUTCOME IS CONTINUING. 320-15-06 - RS GLENOID PLATE EXT CAG +10MM CAGE PEG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 433184 | RS GLENOID PLATE EXT CAG +10MM CAGE PEG | PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED | KWT | EXACTECH, INC. | UNK | UNK | 10885862186706 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |