FDA Adverse Event Malfunction Summary report: N

ELECSYS TROPONIN T HS

MDR report key: 16283311 · Received February 2, 2023

Report

Report Number
1823260-2023-00317
Event Type
Malfunction
Date Received
February 2, 2023
Date of Event
January 12, 2021
Report Date
February 2, 2023
Manufacturer
ROCHE DIAGNOSTICS
Product Code
MMI
PMA / PMN Number
K201441
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE JOURNAL ARTICLE IS "THE UNDERESTIMATED ISSUE OF NON-REPRODUCIBLE CARDIAC TROPONIN I AND T RESULTS: CASE SERIES AND SYSTEMATIC REVIEW OF THE LITERATURE" BY JULIEN FAVRESSE*, JEAN-LOUIS BAYART, DAMIEN GRUSON, SERGIO BERNARDINI, ALDO CLERICO, AND MARCO PERRONE. CLIN CHEM LAB MED 2021; 59(7): 1201¿1211.

Description of Event or Problem · 0

WE RECEIVED AN ALLEGATION OF AN UNDERESTIMATION OF NON-REPRODUCIBLE CARDIAC TROPONIN T RESULTS FROM A JOURNAL ARTICLE. IN MAY-2019: CASE REPORT PATIENT 1- THE INITIAL RESULT OF THE INITIAL SAMPLE WAS 102.0 NG/L. THE RESULT OF A 2ND SAMPLE AFTER 5 HOURS WAS 15.1 NG/L. THE REPEAT RESULT OF THE INITIAL SAMPLE WAS 14.4 NG/L. THE REPEAT RESULT OF THE 2ND SAMPLE WAS 14.0 NG/L. IN DEC-2019: CASE REPORT PATIENT 2- THE INITIAL RESULT OF THE INITIAL SAMPLE WAS 84.0 NG/L. THE RESULT OF A 2ND SAMPLE WAS 6.8 NG/L. THE REPEAT RESULT OF THE INITIAL SAMPLE WAS 5.3 NG/L. CASE REPORT PATIENT 3- THE INITIAL RESULT OF THE INITIAL SAMPLE WAS 393.0 NG/L. THE RESULT OF A 2ND SAMPLE AFTER 3 HOURS WAS 10.3 NG/L. THE REPEAT RESULT OF THE INITIAL SAMPLE WAS 12.0 NG/L. THE REPEAT RESULT OF THE 2ND SAMPLE WAS 13.0 NG/L. IN SEP-2020: CASE REPORT PATIENT 4 - THE INITIAL RESULT OF THE INITIAL SAMPLE WAS 42.3 NG/L. THE RESULT OF A 2ND SAMPLE AFTER 4 HOURS WAS 10.8 NG/L. THE REPEAT RESULT OF THE INITIAL SAMPLE WAS 11.8 NG/L. THE REPEAT RESULT OF THE 2ND SAMPLE WAS 12.0 NG/L. THE ARTICLE IS ATTACHED TO THIS REPORT. THE REAGENT LOT NUMBER AND THE EXPIRATION DATE USED WERE NOT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
372183 ELECSYS TROPONIN T HS IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ROCHE DIAGNOSTICS NA ASKU

Patients

Seq Age Sex Outcome Treatment
1 80 YR Female CASE REPORT PATIENT 4: CICLOSPORIN