FDA Adverse Event Malfunction Summary report: N

CENTURION

MDR report key: 162830 · Received April 9, 1998

Report

Report Number
1824619-1998-90001
Event Type
Malfunction
Date Received
April 9, 1998
Date of Event
March 8, 1998
Report Date
March 16, 1998
Manufacturer
TRI-STATE HOSPITAL SUPPLY CORP.
Product Code
EFT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

ORAL SWAB WAS USED ON PT'S THROAT. FOAM (SPONGE) CAME OFF STICK AND WAS ASPIRATED BY PT. PT STROKE AFTER THIS EVENT. TESTED OTHER ORAL SWABS. MOISTURE WOULD LOOSEN FOAM, AND IT WOULD SLIP OFF. ALL CENTURION MC200 MOUTH CARE KITS WERE PULLED FROM FLOORS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTURION MOUTH CARE KIT EFT TRI-STATE HOSPITAL SUPPLY CORP. * 01268

Patients

Seq Age Sex Outcome Treatment
1 78 YR Life Threatening| R