FDA Adverse Event Injury Summary report: N

ENDURON 10D 58 OR 70ODX28ID

MDR report key: 1627938 · Received March 9, 2010

Report

Report Number
1818910-2010-01203
Event Type
Injury
Date Received
March 9, 2010
Date of Event
February 10, 2010
Report Date
February 10, 2010
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LPH
PMA / PMN Number
K944538
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION SINCE ITS RELEASE FOR DISTRIBUTION IN 06/1997. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFO BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Description of Event or Problem · 1

PT WAS REVISED TO ADDRESS POLY WEAR OF THE LINER AND OSTEOLYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDURON 10D 58 OR 70ODX28ID 87HRY LPH DEPUY ORTHOPAEDICS, INC. NA RT6BF1

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention