FDA Adverse Event Malfunction Summary report: N

COYOTE ES

MDR report key: 16277356 · Received February 1, 2023

Report

Report Number
2124215-2023-03600
Event Type
Malfunction
Date Received
February 1, 2023
Date of Event
January 26, 2023
Report Date
February 1, 2023
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
LIT
UDI-DI
08714729767169
PMA / PMN Number
K093636
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Description of Event or Problem · 0

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT PRESENTED WITH LOWER EXTREMITY ARTERIAL DISEASE AND UNDERWENT ENDOVASCULAR THERAPY. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED RIGHT ANTERIOR TIBIAL ARTERY. VASCULAR ACCESS WAS OBTAINED USING THE IPSILATERAL ANTEGRADE APPROACH. AFTER INSERTING A NON-BOSTON SCIENTIFIC (BSC) SHEATH AND PLACING A NON-BSC GUIDE CATHETER, A JUPITER GUIDEWIRE AND A NON-BSC MICRO CATHETER WERE USED TO CROSS THE LESION. A NON-BSC BALLOON CATHETER WAS INITIALLY ADVANCED FOR DILATATION BUT FAILED TO CROSS THE LESION. A 1.2MMX15MMX142CM EMERGE BALLOON CATHETER WAS ADVANCED BUT RUPTURED AT THE DISTAL TO ANTERIOR PART ON THE THIRD OR FOURTH INFLATION. THE SAME EMERGE BALLOON WAS USED TO DILATE THE LESION AGAIN FOLLOWED BY THE SAME INITIALLY USED NON-BSC BALLOON. THE PHYSICIAN DECIDED TO USE A BIGGER SIZE BALLOON AND ADVANCED A 1.5MM X 20MM X 143CM COYOTE ES BALLOON CATHETER. HOWEVER, DURING THE SECOND INFLATION AT 12 ATMOSPHERES, THE BALLOON RUPTURED. THE BALLOON WAS REPLACED WITH ANOTHER OF SAME DEVICE FOLLOWED BY USING ANOTHER NON-BSC BALLOON COMPLETING THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
998267 COYOTE ES CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC CORPORATION 24691 0030073853 08714729767169

Patients

Seq Age Sex Outcome Treatment
1 Unknown GUIDE CATHETER: MACH 1 6FR ST 55 CM| GUIDEWIRE: JUPITER FC, GLADIUS| INFLATION DEVICE: MERIT| INTRODUCER SHEATH: HALO ONE 6FR| MICRO CATHETER: PROMINENT